UT8805E Benchtop Digital Multimeter
37
2.8.2 Limit Operation
The limit operation can indicate the over-range signal base on the set high/low limit, which includes the
measurements of DC/AC voltage, DC/AC current, resistance, frequency, cycle, capacitance and
temperature.
Press
→ [Limit] → [On] to enter the interface of Figure 2-43.
Figure 2-43 Interface of limit operating
Menu of limit operation function
Function menu
Options
Description
Limit
On/Off
Turn the limit operation on/off
Low
Set low limit
High
Set High limit
Beeper
On/Off
When it is on, beeper beeps as the reading exceeds limit.
Clear
Clear current data and restart the statistic
Done
Save current settings and return to previous menu
Low Limit
Current set low limit
High Limit
Current set high limit
Status
Indicate the status of limit operation (Pass/Fail)
Low Failures
The number of low limit exceeding failures
High Failures
The number of high limit exceeding failures
1) Limit setting method
Select [High] or [Low], switch digits by left/right direction buttons, and modify digits by up/down
direction buttons.
2) Unit of limit
The unit of limit is decided by current measuring function.
3) Over-range indicator
The font of primary measurement will change to red from black and beeper will beep if the reading is
exceeding the high/low limit.
4) Range of limit operation
The set high limit should always higher than the low limit.
The high/low limits are saved in volatile memory, and will be auto reset to default value after power
off.