Model:
UT61B
MANUAL
The measured value shows on the display.
4.
If you need to measure duty cycle, press Hz% button to select % measurement.
Note:
• Input amplitude: When 10Hz~10MHz: 300mV < a < 30Vrms
H. Temperature measurement (see fi gure 9)
The included K type point contact temperature probe can only be used in the measurement
below 230°C. To avoid harm yourself, do not input voltages higher than DC 60V or AC
30Vrms.
To measure temperature, connect the meter as follows:
1. Set the rotary switch to °C °F
2. Insert the temperature probe into the input terminal as shown on fi gure 9
3. Place the temperature probe to the object being measured. After a few seconds, the
measured value shows on the display.
4. Press BLUE button to switch between °C and °F temperature.
Note:
• To avoid measurement error, especially for low temperature measurement, the operat-
ing temperature should not exceed 18°C ~28°C.
D
ATA
O
UTPUTTING
• Press and hold REL
for over 2 seconds to enter or exit RS232C or USB mode.
• The sleep mode feature will be disabled after entering RS232C or USB mode, the
symbol on the LCD display will disappear.
• If the meter is carrying out HOLD, MAX/MIN or REL measurement, the LCD will display
the corresponding readings, but the interface output will still be the random value of
the present input terminals measurement.
S
LEEP
M
ODE
To preserve battery life, the meter automatically turns off if you do not turn the rotary
switch or press any button for around 15 minutes. The meter can be activated by turning
the rotary switch or pressing any button. To disable the Sleep Mode function, press BLUE
button while turning on the meter.
14