13.1 Auto Setting
Automatic settings will choose appropriate time base scales, amplitude scales and triggering parameters according to the
input signal so that the waveform automatically properly displays on the screen. Press the AUTO key to enable automatic
settings.
Automatic setting applies only to the following conditions:
①
Automatic setting is suitable for setting up simple single frequency signals, and not for complex combination waves.
②
The measured signal frequency is not less than 20Hz, the amplitude is not less than 20mVpp, and the square wave duty
cycle is more than 5%.
③
The automatic setting is only for the open channels and not for the closed channels.
13.2 Run / Stop
Use the RUN/STOP key on the front panel for control. When the key is pressed and the green light is on, it indicates the
RUN state, and if the red light is on after the key press, it is the STOP state. In the running state, the oscilloscope is
continuously acquiring waveforms and the upper part of the screen shows “AUTO”; in the stop state, the oscilloscope
stops the acquisition and the upper part of the screen shows “STOP”. Press the RUN/STOP key to switch the waveform
sampling between the run and stop states.
13.3 Clear
Press the CLEAR button on the front panel of the oscilloscope to clear the loaded REF waveform on the screen.
13.4 Factory Setting
By pressing the DEFAULT button, you can quickly restore to factory settings. UPO2000E series oscilloscope factory
settings are as below:
Chapter 13 Additional Function Keys
Factory Setting
1V/DIV
0 (vertical midpoint)
DC
Off
Coarse tuning
1×
Off
Off
Off
Off
Off
1μs/div
0 (Horizontal midpoint)
80.00ns
Edge
CH1
Rising
DC
Automatic
Vector
YT
50ms
40%
40%
Function
CH1
Vertical Displacement
Coupling
Bandwidth Limitation
VOLTS/DIV
Probe
Invert
Bias Voltage
CH2, CH3, CH4
MATH, REF
Extended Window
Horizontal Time Base
Horizontal Displacement
Hold-off Time
Trigger Type
Source
Slope Type
Coupling Mode
Trigger Mode
Type
Format
Duration
Grid Brightness
Waveform Brightness
System
Vertical System
Horizontal System
Trigger System
Display
57
Summary of Contents for UPO2000E Series
Page 1: ...UPO2000E Series Ultra Phosphor Oscilloscope User Manual...
Page 66: ......