Service Manual MSO/UPO3000E Serials products
Instruments.uni-trend.com
24
/
27
•
electricity failure
•
One or more defective load regulator points
The display is blank or there are streaks in the display
• LCD screen or video circuit failure.
Required equipment
•
Digital voltmeter for checking mains voltage.
•
Anti-static working environment.
Troubleshooting flowchart
Follow the troubleshooting flowchart to determine the action for the fault condition.
Start
Does the power bu tton
turn red?
Press the power
button
Confirm
:
Is the power cor d connected to the
oscil loscope?
Is the power cor d connected to the AC
outl et?
Does the AC outlet have AC power?
Replace the power cord
no
yes
yes
Connect the power
cord
Does the power bu tton
turn green?
Does the fan spin?
yes
After 2 mi nutes, is
the display normal?
yes
Does the oscilloscope pass
the power-on se lf-test?
yes
Stop
yes
no
no
Power supply ch ec k
of the power modu le
Check the ground
resistance of each
secondary power supply
in the board
no
no
Back to UNI
-
T Servi ce
Center
Does the power bu tton
turn green?
no
Figure 2: Main troubleshooting tree