Service Manual MSO/UPO3000E Serials products
Instruments.uni-trend.com
15
/
27
Remove the power module
The following procedure describes the removal and replacement of a power module.
prerequisites:
•
To prevent electrostatic damage to components while working on the instrument, wear a properly grounded anti-
static wrist and foot strap, and use an anti-static mat in a tested anti-static environment.
•
Remove the rear cover.
•
Remove the rear assembly.
step:
1.
Use a PH1 Phillips-tipped screwdriver to remove the two screws from the rear chassis frame.
2.
Use a PH1 Phillips-tip screwdriver to remove two more screws from the rear assembly and remove the power supply
cover.