OPERATING INSTRUCTIONS
30
USER MANUAL — 74-36105-ENG R5
Short power cord (optional)
The power cord length can be adjusted to meet the needs and condition of a patient in a specific environment. The
length of the short power cord is of 1 m / 40", but can be adjusted to a longer length if required. The label fixed on
the power cord always needs to be visible to ensure that the minimum length allowed for this option is respected;
do not remove the label from the power cord.
WARNING
•
Verification, and maintenance of bed, and power cord adjustment shall be performed by qualified and
trained personnel. Electrical or electronic misconnection could result in shock or fire hazards.
• Always ensure that usage of a power cord with reduced length is restricted only to patients with an
evaluated medical condition that requires extreme protection measures. Plug the power cord to a hospital
grade power outlet without the usage of an extension cord.
• The use of the removable patient control option is not recommended if option of short power cord applies
to a patient with an evaluated medical condition that requires extreme protection measures.
CPR release (optional)
CPR release handles are located on each side of the bed frame, under the backrest section. Usage shall be
limited to emergency situations only. An over usage could lead to mechanical degradation.
Operation:
1.
To activate, pull and hold the CPR release handle to flatten all sections of patient sleep surface.
2.
Release handle once surface has reached flat position. If released before surface is completely flatten, all
sections will stop retracting.
WARNING
• CPR release option usage shall be restricted to emergencies. To avoid injuries or product damages, ensure
total clearance around product prior to usage.
ATTENTION
• Usage shall be limited to emergency situations only. An over usage could lead to mechanical degradation.
•
If bed operates on battery, the thigh section will not reach flat postion.
IMPORTANT
Thigh section will not reach flat position if master lockout of thigh section or, a movement function of any control
is activated during motion.
Nurse call connector (optional)
If the bed is equipped with the optional nurse call system, the connector selected when ordered will be located
at the head end of the bed, on the frame. The connector is only designed for nurse call function and room light
control. Communication will be sent through system for a Nurse call initiated by the patient and/or for a bed exit
control audible signal. Ensure the plug is properly inserted into connector.
!
!
!