USER MANUAL — 74-36105-ENG R5
25
OPERATING INSTRUCTIONS
IMPORTANT
Read and understand all information covered in this manual before operating the equipment. Users shall be
knowledgeable about inherent hazards related to use of electric beds.
Power up and modes of operation
If a power failure occurs, bed will automatically fall in sleep mode and initiate the battery-powered mode.
If the bed operates on batteries, the patient and caregiver controls for bed-motion will remain functional. A lockout
on siderail motion control will be saved.
Powered mode:
Powered mode is activated if bed is power supplied or if power returns after failure.
All bed functionalities will operate if bed is power supplied. If battery level requires
recharge, the battery pictogram with integrated LED will be flashing green.
Screen sleep mode: Activated after 1 minute of bed inactivity; screen brightness will
decrease by half. After 2 minutes of bed inactivity, screen will turn off completely and
get in sleep mode. If a screen saver is activated, the information will be displayed
while in sleep mode. To get out of sleep mode, press on any button associated with
the screen.
Battery powered mode: Battery powered mode is activated if power failure occurs or if bed power cord is
unplugged from wall outlet ; it will first fall in sleep mode. Under the battery-powered
mode, an armed bed exit control system will send a signal to the nurse call
communication system to inform of battery-powered mode; no local sound signal
will be heard. Only the UP/DOWN press buttons, the Trendelenburg and reverse
Trendelenburg press buttons for bed motion and the motion lockout will operate.
Will not operate: 30° backrest positioning and vascular positionning, bed exit control or
scale system, display screen and total lockout. The battery charge level is accessible
on the footboard control or on the optional back-up control on frame.
Low level audible signal: Two LED sections of battery pictogram lighted. When a
movement function is activated, functionalities will operate with a low battery level
audible signal.
Recharge audible signal: One red LED section of battery pictogram lighted. When
a movement function is activated, functionalities will not operate and there will be a
constant audible signal.
Summary of Contents for ooksnow FL36
Page 1: ...2 0 1 7 1 0 7 4 3 6 1 0 5 E N G R 5 User manual FL36 MH...
Page 2: ......
Page 3: ...2 0 1 7 1 0 7 4 3 6 1 0 5 E N G R 5 User manual FL36 MH...
Page 46: ......
Page 58: ......
Page 74: ......