INTRODUCTION
USER MANUAL — 74-36105-ENG R5
9
Environmental conditions
Use
ambient temperature
relative humidity
atmospheric pressure
50 °F to 104 °F
5 to 95% without
condensation
500 to 1060 hPa
10 °C to 40 °C
5 to 95% without
condensation
500 to 1060 hPa
Transport and storage
ambient temperature
relative humidity
atmospheric pressure
-40 °F to 158 °F
10% to 100%
500 hPa to 1060 hPa
-40 °C to 70 °C
10% to 100%
500 hPa to 1060 hPa
Maximum altitude
13,123 ft
4000 m
Electrical requirements
Electrical and mechanical requirements
comply with standards: IEC60601-1, CAN/CSA C22.2
60601-1, UL60601-1, IEC60601-1-2, IEC60601-2-52,
CAN/CSA C22.2 60601-2-52
Degree of protection
IPX4
Degree of protection against electrical shocks
Class I, Type B
Scale accuracy
0.5% with a difference of maximum 1.1 lb / 0.5 kg
Voltage
120 V AC, 60 Hz / 220-240 V AC, 50-60 Hz
Amperage (includes power outlet and powered system)
120 V AC: Bed 4.25 A / Auxiliary outlets (optional) 5 A /
Total = 9.25 A
220-240 V AC: Bed 2.15 A
Current leakage
<300 micro-amperes
Battery type
sealed rechargeable lead acid battery 1.2 Ah, 2 x 12 V
Actuator duty cycle
2 minutes On, 18 minutes Off
Auxiliary power outlet (with 120V AC model only)
120 V AC / 5 A
USB type A charger
5 V DC / 3.1 A