SK00-4802-DI-002-04
No.16.10.20
vii
Be sure to supply water during the operation. The required amount of
cooling water is as follows:
・
Amount of cooling water
ULK-04A/D-11(0.55kW) : 1.0 L/min or more
ULK-04A/D-31(0.73kW) : 1.0 L/min or more
ULK-06A/D-11(0.9kW) : 1.0 L/min or more
ULK-06A/D-31(1.2kW) : 1.5 L/min or more
ULK-10A/D-11(2.0kW) : 2.5 L/min or more
ULK-10A/D-31(2.4kW) : 3.0 L/min or more
ULK-14A/D-11(2.25kW) : 2.5 L/min or more
ULK-14A/D-31(2.4kW) : 3.5 L/min or more
・
Cooling water pressure: 0.1~ 0.3MPaG
(Gauge pressure)
・
Cooling water temperature: 19 ~ 25
℃
We recommend using water with the low amount of impurities (such as
industrial water shown in the table below) for cooling water of this product.
Water stain, such as calcium carbonate, accumulates in the cooling water
system of this product depending on the water quality, which reduces the
cooling efficiency. In addition, chlorine ions may corrode from the inner
wall, causing cooling water leakage. Note that in such a case, the repair
will be charged.
Established by: Japan Industrial Water Association (Industrial Water Quality Criteria
Establishment Committee)
Turbidity
(mg/L)
pH
-
Alkali level
CaCO
2
mg/L
Hardness
Mg, CaCO
3
mg/L
Evaporation
residue
mg/L
Chlorine
ion
Cl
-1
Mg/L
Iron
Fe
mg/L
Manganese
Mn
mg/L
20
or less
6.5-8.0
75
or less 120
or less
250
or less
80
or less
0.3
or less
0.2
or less
[Reference] Supplied Standard Water in Japanese Industrial Water Service
Power Cable <Removal>
Before removal, be sure to disconnect it from the power supply.
The pump keeps its temperature extremely high while the pump is running
or for a while after it stops. If someone's body comes into contact with it,
the person may get burned. Supply the cooling water until the pump
temperature becomes low.
CAUTION
CAUTION
WARNING