background image

 

 

7

   

3.4  Treatment of Trapped Oil 

The TM201/401 is designed to discharge the trapped oil to outside through 

the oil return port.    Recover the discharged oil in a container and return it to 
the pump when the pump is operated under no load or at stop. 

  However, do not return the oil to the pump if the oiliness of the recovered 

oil is lower than the new oil used in the pump or if the oil contains water, dust 
or the like or if chemicals like acid is used. 

  It can be a cause of trouble of the pump. 

CAUTIONS: 

 

If you do not dispose of the trapped oil, oil will stay between the 

fitter element and the oil mist trap and oil mist will come out. 

 

It is recommended to replace a filter element once immersed in oil 

because clogging tends to occur. 

 

Do not connect the oil return port and the pressure monitor mounting 

port by means of a vinyl hose or the like.  Oil mist will not be trapped. 

 

Summary of Contents for TM201

Page 1: ...SK00 6740 DI 002 03 No 170821 INSTRUCTION MANUAL OIL MIST TRAP MODEL TM201 TM401 Components division ULVAC Inc Read this manual before operation and keep it at your hand for immediate reference...

Page 2: ...il 7 4 INSPECTION 8 4 1 Schedule Inspection 8 4 2 Replacing the Filter Element 8 4 2 1 Removing the filter element 8 4 2 2 Reassembling the filter element 10 5 PRECAUTIONS 11 5 1 Clogging of Filter El...

Page 3: ...Fig 1 Dimensional drawing TM201 3 Fig 2 Dimensional drawing TM401 4 Fig 3 Schematic diagram 5 Fig 4 TM201 TM401 mounted in position 6 Fig 5 Exploded view of filter element 9 Table 1 Specifications 2 T...

Page 4: ...oil mist trap read this manual carefully for optimum performance and operate it correctly Note that incorrect operation may cause trouble of the oil mist trap Also keep this manual in your file so th...

Page 5: ...l rise at start up of the pump or during pumping operation which can be dangerous If the pressure in the pump rises above 0 03 MPa 0 3 kg cm2 gauge pressure the pressure relief valve will be activated...

Page 6: ...3...

Page 7: ...4...

Page 8: ...h fine particles of oil blown up These fine particles of oil oil mist are trapped by the filter element when they pass through it and the condensed large oil particles gather between the oil mist trap...

Page 9: ...r local ULVAC organization or representative 3 2 Mounting on Pump Refer to Fig 4 To mount the oil mist trap on the pump proceed as follows 1 Verify that the exhaust port O ring groove and oil mist tra...

Page 10: ...than the new oil used in the pump or if the oil contains water dust or the like or if chemicals like acid is used It can be a cause of trouble of the pump CAUTIONS If you do not dispose of the trapped...

Page 11: ...place it as a rule 4 2 1 Remove the filter element 1 Stop the pump or put the pump under no load operating condition and loosen the Allen screw M8 that fixes the oil mist trap to the pump to remove th...

Page 12: ...9...

Page 13: ...ing the spacer nut 5 Mount the parts of pressure relief valve CAUTIONS Tighten the hex nut M10 until it comes into contact with the spacer nut The pressure relief valve is activated at a pressure belo...

Page 14: ...ent temperature it will be deposited on the fiber of the filter element leading to clogging When reaction products are generated 5 2 Pressure Relief Valve This device is designed to operate when the p...

Page 15: ...rent one is recommended 5 4 When using Oil Mist Trap for a Pump that exceeds Maximum Throughput Do not operate the pump continuously at a high intake pressure of 6500 Pa or more when the TM201 is used...

Page 16: ...bient temperature 10 to 40 b Suction gas Dry air or dry nitrogen 0 to 40 c Operation in conformity with instruction manuals 3 Troubles caused by the following are out of warranty a Acts of God and for...

Page 17: ...t No Description Material Q ty Remarks 1 Filter element 1 set TM201 TM201E TM401 TM401E 2 O ring G145 NBR 1 JIS B 2401 3 Seal washer M10 NBR 1 The exhaust port of the TM201 and TM401 are provided with...

Reviews: