Output Signal 

  The capacitance manometer output appears at pins 2 and 12 of the 

manometer connector. 

The output signal is linear and the output is 10 volts on the full scale. 

 

Example

Gauge with CCMT-10D,measureing signal Vout=6V 

P=(6/10)×10×(101325/760=133.3224

・・・

=799.9344 (Pa) 

 
 

Zero Adjustment 

We recommend performing a zero adjustment, when the gauge is operated 

for the first time. 

Outputs from this unit vary slightly with ambient temperature variation. 

Model CCMT-D: 0.005%/°C of the full scale 

Also the zero point may drift due to power variation immediately after 

power is turned on or deterioration or mounting orientation or contamination 
of the detector. In that event, make zero adjustment. 
 
1. Measure the stability of the indicated value after waiting for more than two 

hours of turning on power. If automatic zero point correction is being 
made by the controller, reset it. 

 
2. Make zero adjustment after pumping down the manometer to below 

pressure. 

 

CCMT-1000D  

  6.65     Pa or less 

CCMT- 100D  

  6.65 x10

-1 

Pa or less 

CCMT-  10D  

  6.65 x10

-2 

Pa or less 

CCMT-   1D  

  6.65 x10

-3

 Pa or less 

 
3. Push the <ZERO> button with a pin (max. ø1.1 mm). The zero adjustment 

runs automatically. The <RUN> LED flashes until the adjustment is 
completed. 
The <RUN> LED flashes if the signal output is negative. 

 
4.

 LED will flash when the range of zero adjustment rise the 

range. 

 

Fig.2 <ZERO> button location 

 

!

 

CAUTION

 

Operating conditions 

When you do zero point adjustment, please use the

 

  

attached 

pin(max.φ1.1mm)

 
 

The influence to the pressure indication by mounting 
orientation 

Due to the influence of gravity, the mounting orientation 

and the barrier membrane displacement is changed. Table below 
states the impact of the vertical direction with respect to 
the horizontal direction of brand new capacitance manometer.

 

Model 

Estimated deviation range 

CCMT-1000D 

Around   2mV

26.7Pa

 

CCMT-100D 

Around  10mV

13.3Pa

 

CCMT-10D 

Around  50mV

 6.7Pa

 

CCMT-1D 

Around 300mV

 4.0Pa

 

 

 

Warranty 

This product was shipped after rigid company inspection.  However, in 

case any failure occurs under ULVAC’s responsibility, such as defect in 
manufacturing and damage during transportation, Buyer shall inform ULVAC, 
Inc. or the local ULVAC representatives.    ULVAC will repair or exchange it 
at free of charge. 
 
Warrantable Items: Capacitance Manometer delivered to the customer 
 
Duration of guarantee: One (1) year after shipping date from ULVAC 
 
Warrantee scope 

1)

 

Domestic business in Japan: Product, which has damage, caused by a 
failure on delivery. 

2)

 

Direct export transaction: Product, which has damage, caused by a failure 
on delivery.   The warrantee scope shall confirm to the new 
INCOTERMS. 

3)

 

Products not satisfying meet the standard specifications although the 
product is used under the normal service conditions such as temperature 
range and power etc. 

 
Response procedure 

1)

 

Domestic business in Japan: ULVAC send a replacement or Buyer return 
the defective items to ULVAC, Inc. or to the local ULVAC representatives 
for repair.    If field service is required, Buyer shall ask ULVAC, Inc. or 
the local ULVAC representatives. 

2)

 

Direct export transaction: ULVAC send a replacement or Buyer return the 
defective items to ULVAC, Inc. or to the local ULVAC representatives 
for repair.    Return charge shall be paid by Buyer. 

 
 
 

Disclaimer 

1)

 

Failure occurred after expiration of warranty period 

2)

 

Failure caused by force majeure, such as fire, storm and flood damage, 
earthquake, lightning strike, war etc 

3)

 

Failure occurred due to carelessness handling or faulty usage 

4)

 

Products remodeled, disassembled or repaired without ULVAC’s 
acceptance 

5)

 

Failure occurred under abnormal environment, such as intense 
electromagnetic field, radiation, high-temperature, high-humidity, 
flammable gases, corrosive gases, dust etc. 

6)

 

Failure occurred by noise 

7)

 

Product deficiency or secondary damnification occurred to Buyer, from 
law suit to ULVAC by third party for patent infringement. 

8)

 

Sensor head being used (expiration of life, measurement error, etc.) 

9)

 

Sensor head cable being used (cable burnout due to improper installation, 
poor contact, etc.) 

 
Others 

1)

 

In case, special agreement or memorandum for specifications is made 
individually, the descriptions are prior to this article  “13 Product 
Warranty”. 

2)

 

Buyer shall inform ULVAC when this product is exported out of Japan.   
In the meantime, Buyer shall take necessary procedures according to 
Foreign Exchange and Foreign Trade Law. 

3)

 

As for the question and consultation, Buyer shall check the model and 
serial number and ask the local representative or ULVAC, Inc. 

4)

 

The content of this document is subject to change without notice in future. 

 
 
 

Troubleshooting 

Symptom: This unit remains at about 10.5 volts and does not change when 

pressure has changed. 

 

Pressure is still higher than the measurement range. 

→Normal

 

 

Considered the possibility of breakdown. 

→To be inspected and repaired by ULVAC.

 

 
Symptom

Manometer output value is obviously unnatural. 

 

Contamination or deterioration of the manometer 

Replace the manometer 

 

Temperature in the place where the manometer is installed deviates 

greatly from 25°C or temperature variation is high. 

→Change the install location. (Select a clean, well ventilated pla

ce not 

blown with direct draft.) 

 

Zero point is shifted. 

→Make zero adjustment again.

 

 
Symptom

Manometer output does not indicate a constant value. 

 

Pressure is actually changing. 

→Normal

 

 

Contamination or deterioration of this unit. 

→Replace the 

manometer with a new one. 

 

There is a leak in this unit or area where this unit is installed. 

→Normal

 

 

There is a leak in this unit or area where this unit is installed. 

→Leak in the manometer .

 

 Leak in others .Stop the leak. 

 

The is a vibration 

Install the manometer to nor vibrate. 

 

Symptom

The zero adjustment function does not work.

 

 

Pressure of chamber is high. 

 Pump down the chamber pressure under of the indicated 

pressure (Refer to Zero Adjustment).  

 

RUN LED flashes after the zero point adjustment 

It is exceeding the zero point adjustment range.  
Replace the manometer. 

 
 

Network 

ULVAC,Inc: http://www.ulvac.co.jp/eng/index.html 
Service Centers: http://www.ulvac.co.jp/eng/support/service/index.html 
Sales Offices: http://www.ulvac.co.jp/eng/support/sales_office/index.html 

 

 

Contamination Certificate 

This form is a contamination certificate to be presented to ULVAC before requesting ULVAC to make 
repair/inspection of ULVAC components. 
Please fill in this certificate and forward it to your local ULVAC representative along with 
component(s)when requesting repair or others. 
Regarding components used in toxic gas or components deposited with reactive products, contact your 
local ULVAC representative before shipping component(s). 
Name of product: 
Model name    : 
Serial No.      : 
Application     : 
Request to ULVAC ___________________________________ 
Trouble symptom  ___________________________________ 
Note            ____________________________________ 
Contaminant (Check 

□ whichever applicable)

 

□ We certify that the product above is not contaminated with toxic substance(s).

 

□ The product above is contaminated with the following toxic substance(s).

 

 

Name of contaminant (molecular form) 

Characteristics 

 

 

 

 

 

 

 

 

 

 

 
To: ULVAC, Inc. 

 

 

 

 

 

     ,       

Name of personnel in charge: ____________   
Division                : _______________ 
Name of your company      : ______________ 
Phone                  : ______________ 
Fax 

         : _______________ 

E-mail   

       

: _______________ 

 
*   

The responsibility for accidents that may be caused by contaminant in transit to 

ULVAC lies with the user (shipper). So be careful in packing. ULVAC may decline your request for 
repair depending on the type of contaminant and degree of contamination and return it to you.   

 

By ULVAC, Inc. 
MSDS request: yes/no 

Signed by 

 

ULVAC job No. 

 
 

China RoHS 

 

 

This mark is applied to the electronic information product sold in the 
People's Republic of China.    The figure at the center of the mark is 
the validity date of environmental protection. This product does not 
influence the environment, the human body and the property during 
the period reckoning the manufacturing date as long as the cautions 
for safe use regarding the products are observed.   
*The environmental protection validity date is not the product 

warranty period. 

 

Table1. Making format for names and contents of hazardous substances or elements 

Name of parts 

Hazardous substances or elements 

Pb 

Hg 

Cd 

Cr

6+

 

PBB 

PBDE 

Printed Circuit Board 

× 

 

 

 

 

 

Frame 

 

 

 

 

 

 

Connector 

× 

 

 

 

 

 

Label 

 

 

 

 

 

 

 

 

 

 

 

 

 

○: indicating that content of the 

hazardous substance or element in all homogeneous 

materials of the part does not exceed the requirements for concentration limits 
specified by SJ/T11363-2006. 

×: indicating that content of the hazardous substance or element in, at least one kind 

of, homogeneous materials of the part exceeds the requirements for concentration 
limits specified by SJ/T11363-2006. Producer may further explain the technical 
excuse to the items marked with “X” perspecific conditions here. 

 
 
 
 

 

ULVAC, Inc. 

Components Division, 

http://www.ulvac.co.jp/

 

0

10

20

30

40

50

60

70

80

90

100

0

1

2

3

4

5

6

7

8

9

10

A

ga

in

st sc

ale(

%

)

Capacitance manometer output(V)

P=(Vout/10) * P(F.S.)

 

ZERO

RUN

 

Outside Dimensional Drawing 

113

69

4

(28

.4

φ55

Reviews: