Prior to Operation

 

Upon receipt of this detector, make sure that it is the correct model you ordered 
and that it is not damaged in transit. 
 

!

 

WARNING

 

Read this manual before installation, operation, 
inspection or maintenance and fully understand safety 
cautions, specifications and operating procedure of 
this instrument so that the unit can be used for a long 
time. 

!

 

NOTICE

 

No part of this manual may be reproduced for use by a 
third party without written consent by ULVAC. 

 

Safety Denotations

 

!

   

WARNING

 

The following denotations are used throughout this 
manual to call operator's attention to safety.    The 
denotations are classified as follows. 

 

!

 

DANGER

 

Failure to comply with DANGER involves possibility 
of impending loss of life or serious personal injury. 

!

 

WARNING

 

Failure to comply with WARNING involves the 
possibility of loss of life or serious personal injury. 

!

 

CAUTION

 

Failure to comply with CAUTION can lead to a 
medium degree of personal injury or serious damage 
to the equipment. 

!

 

NOTICE

 

Failure to comply with NOTICE can lead to damage 
to the equipment or malfunction. 

 

Safety Cautions

 

For safe use of this Ceramic Capacitance Manometer, carefully read this 

manual and comply with the warnings and cautions given in the manual. 

!

 

WARNING 

Turn OFF power. 

If the unit fails, immediately turn OFF the power. 
Use of a failed unit may cause fire or electric shock. For 
repair, contact your local ULVAC representative or 
ULVAC,Inc.. 

!

 

WARNING 

Turn OFF power. 

Before touching the power terminal of this unit or if there is 
a possibility of touching it, turn OFF the power to this unit. 
Some of the power terminals are energized with a line 
voltage of +14

30 VDC. Contact with the terminal will 

cause an electric shock. (When GM-1000, GM-2001 or GM-
2002 is used, the terminal is also energized with -15 volts.) 

!

 

WARNING

 

Operating environment 

Do not connect this unit, except the VCR joint type, to a test 
object of which pressure is in excess of atmospheric 
pressure.    If the pressure in this unit exceeds atmospheric 
pressure, this unit will pop out from the connection, causing 
damage to the surrounding, including injury to human body. 
If the pressure in this unit exceeds the maximum operating 
pressure, this unit may be damaged or fail. If it is likely that 
the pressure exceeds the maximum operating pressure, 
provide an isolation valve so that the pressure in this unit 
does not exceed the maximum operating pressure. 

!

 

CAUTION

 

Don't disassemble. 

Do not disassemble this unit. 

!

 

CAUTION

 

Don't modify. 

Do not modify this unit. If it is modified, its functions are 
not warranted.    Also fire or electric shock may result. 

!

 

CAUTION

 

Check connection. 

Ground the power ground and frame ground. 

!

 

CAUTION

 

Check connection. 

Prior to turning ON the power, make sure that there is no 
error in wiring. Connecting incorrect power may cause 
damage to this unit or fire. 

!

 

CAUTION

 

Operating conditions 

When you do zero point adjustment, please use the attached 
p

in(max.φ1.1mm)

!

 

CAUTION

 

Check line voltage. 

Prior to turning ON the power, make sure that this unit 
operating voltage and the supply voltage are in agreement. 
Connection of incorrect power can cause damage to this unit 
or fire.The supply voltage is +14

30 VDC (pin 7 or 11). 

!

 

CAUTION

 

Operating conditions 

Use this unit under the environment set forth in the 
specification. 

!

 

CAUTION

 

Operating environment 

Do not use this unit in a place where it may be splashed with 
water. If it is splashed with water, failure, earth leakage or 
fire can result. 

!

 

CAUTION

 

Beware of impact. 

Do not give an impact to this unit. 

!

 

CAUTION

 

Repacking for transfer 

If this unit is to be shipped to another site, repack it in the 
same way as on delivery. If it is shipped bare, it may be 
damaged. 

!

 

CAUTION

 

Discarding 

When discarding this unit, comply with your local 
regulations. Especially, if this unit was used in an 
atmosphere that can cause hazards to human body, it must 
be disposed of by a specialist in disposal. Customer shall be 
responsible for the cost relating to disinstallation. 

!

 

CAUTION 

Operating conditions 

When you do zero point adjustment, please use the

 

  

attached pin(max.φ1.1mm).

 

 
 

Feature 

The Model CCMT-D is a capacitance manometer that detects 

displacement of ceramic diaphragm caused by change of gas pressure by 
utilizing electrostatic capacitance. 

The heart of the sensor being a mono block made of ceramics, this 

manometer provides high repeatability and stability and permits high 
accuracy measurement of pressure in the low and medium vacuum regions 
regardless of the type of gas. 
 
 

Specifications 

 
Please refer to the spec chart. 
 
 

Accessories 

Pin(max.φ1.1mm)

 

×1 

Instruction manual      (this manual) 

×1 

 
 

Option 

Display 

ISG1

GM-2001

GM-2002 

Display Cable 

 

Test results certificate 

 

Calibration certificate 

Calibration certificate, JCSS Calibration 
certificate 

 
 

Installation of CCMT-D 

1. Preliminary Operation 

Unpack this unit and check quantities. 

Check components to see if any of them is damaged in transit. 

 
2. Installing this unit 

Install this unit to the gauge port of a vacuum system. 

2.1. Measuring position 

This unit measures static pressure in the position where it is connected. If 
there is gas flow or an outgas source or an intense electron or ion 
generating source, the measurement value will be affected. So select the 
measuring position carefully. 

Note that if this unit is subjected to vibration, heat radiation, high intensity 
magnetic field or intense radiation, correct pressure measurement may not 
be made. 

Be careful that the ambient temperature deviates considerably from the 
calibrating temperature (25°C). 

 
2.2. Installing this unit 

Install this unit in such a manner that this unit mounted opening plane is 
parallel with gas flow. See to it that gas does not enter this unit in the form 
of beam. Avoid use in a place where vibration is at a high level, if 
possible.   

Be careful in selecting the installation place and handling this unit. 

To install this unit, use an O-ring that releases little outgas. 

Use of a material that releases much outgas like rubber pipe or grease for 
connection of this unit can be a cause of error.   

 
 
 
 
 
 

Electrical connection 

Use the D-Sub 15-pin female connector (inch screw #4-40) to supply 

power to this unit and to connect the output signal from this unit. Apply 
power of +14

30 VDC to pin 7 or 11. 

The signal from the capacitance manometer is outputted between pins 2 

and 12. 
 
 
 
 
 
 
 
 
 

Pin No. 

Use 

V+ 

Analog ground 

+14

30V 

11 

+14

30V 

12 

V- 

15 

Frame ground 

 

Fig.1. Capacitance manometer connection diagram 

 

!

 

CAUTION

 

Wiring connection check 
Do not connect the PIN  number  1,3,4,6,8,9,10,13 
and 14. 

 
 

Operation 

1. Turning on power to the manometer will start measurement. 

<RUN> LED lighting: measuring 
<RUN> LED flashing: over pressure or zero adjustment 

2. However, wait for more than 30 minutes after turning on the power until 

the output voltage is stabilized.     

Especially when a manometer for measuring a low pressure range (for 
example, Model CCMT-10D) is used or when an effective digit below 
1/1000 of the full scale is required, start measurement after the 
temperature equilibrium of the manometer is fully established by applying 
power for several hours continuously. 
Do not turn off the power when a sequence of measurements is going on. 
At switch on, the output voltage value shows as -5V,    in order to 
initialize the inner circuit. Therefore, please import the output voltage 
after 2 or more seconds from turning on. 

 
3. This is an absolute pressure type capacitance manometer of which 

measurement value is not affected by the type of gas, but its characteristics 
may change if it is exposed to a chemically active gas or highly absorptive 
gas. 

 
4. The proof-pressure of the capacitance manometer is according to the table 

below: 

CCMT-1000D  

  4.0 x10

+5

 Pa    (absolute pressure) 

CCMT- 100D  

  2.6 x10

+5

Pa 

CCMT-  10D  

  2.6 x10

+5

Pa 

CCMT-   1D  

  2.6 x10

+5

 Pa 

 
5. It is necessary to correct the zero point at a pressure sufficiently lower than 

the measurement pressure before starting measurement. 
If the ultimate pressure of the system in which this manometer is used is 
not sufficiently low to make zero point correction, its specifications may 
not be met.    Refer to “Zero Point Adjustment Function.” 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Specifications 

Model 

CCMT-1000D 

CCMT-100D 

CCMT-10D 

CCMT-1D 

Range full scale 

133kPa  F.S. 

13.3kPa F.S. 

1.33kPa F.S. 

133Pa F.S. 

Lowest reading 

13Pa 

1.3Pa 

0.13Pa 

1.3x10

-2

Pa 

Lowest suggested reading 

66.6Pa 

6.7Pa 

0.67Pa 

6.7x10

-2

Pa 

Accuracy

1)

 

±

0.2% of Reading

  ±

  Temperature effect 

Resolution 

0.003% F.S. 

Temperature effect on zero 

0.005% F.S./

 

0.015% F.S./

 

Temperature effect on span 

0.01%  Reading/

 

Ambient operating temperature 

10

50

 

Storage temperature 

-40 

60

 

Max temperature at flange

2)

 

110

 

Pressure max. (absolute) 

4x10

+5

Pa 

2.6x10

+5

Pa 

Response time 

30msec 

Materials exposed to process media 

Aluminum oxide ceramic

Al

2

O

3

)、

Vacon70

3)

SUS316

glass ceramics solder

AgTiCu hard Solder 

Standards 

CE 

Protection type 

IP30 

Supply voltage 

DC14-30V,1W 

Output signal 

DC0-10V 

input impedance(recommended): >10kohm,   

0.1uF 

Flange 

1/2 inch tube

DN 16 ISO-KF

8VCR 

Weight 

1/2 inch tube:329g

DN 16 ISO-KF:344g

8VCR:386g 

Internal volume 

1/2 inch tube:5.8 cm3

DN 16 ISO-KF:6cm3

8VCR:8.15cm3 

Electrical connection 

15 pole D-Sub. Male (inch screw #4-40) 

1)Non-linearity, hysteresis, repeatability at 25°C ambient operating temperature without temperature effects after 2 hours operation

  after adjustment for 

ZERO, 

 

We recommend regular zero point adjustments before use, depending on the customer's environment. 

2)non operation 
3)28% Ni, 23% Co, 49% Fe 

This manual is for the capacitance manometers of the 

following serial numbers. 

CCMT-1000D    

 

S/N 00001

 

CCMT-100D   

 

  S/N 00001

 

CCMT-10D  

 

    S/N 00001

 

CCMT-1D      

 

  S/N 00001

 

 