6
Screen 3: Setting the temperature
The
Neon Series
has three process temperature settings:
Min temperature:
this is the minimum temperature at which the bath will operate. If
the liquid falls below this temperature, the bath will not operate until the liquid rises to
the min. temperature again. The
Neon Series
automatically starts heating the liquid if it
is below the min. temperature.
Normal temperature:
this is the optimum operating temperature. If the liquid
temperature falls more than 2°C below “normal”, the electrical water heater will switch
on until the liquid reaches the “normal” temperature.
Max temperature:
this is the maximum temperature at which the bath will operate. If it
rises above this temperature it will not operate. To overcome this problem, drain the
tank and refill with cooler liquid.
The
Neon Series
is factory pre-set with the following:
-
Minimum temperature
15°C
-
Normal temperature
40°C
-
Maximum temperature
70°C
Use the “Up” and “Down” buttons to scroll through the numbers. When the correct
number has been selected, use the “Next” button to move to the next character.
Pressing ‘Enter’ saves the current set temperatures and advances to the ‘Set Cycle’
screen.
Pressing ‘Esc’ returns to the ‘Set Date’ screen without saving.