14
Step 3
– Open the filter housing bleed valve on the bottom of one of the housings:
Step 4 –
Dispose of all fluid in the filter housing.
Step 5 –
Repeat the same process with the second filter housing.
Step 6
– Remove each filter housing using the tool provided, by turning the housing
locking ring left/anti-clockwise:
Step 7
– The filter can now be inspected and replaced if necessary.
Step 8
– Repeat the same process for the second filter.
Before returning the isolation valves to their normal working positions make sure that
filter housings are replaced and re-tightened.
Return the two isolation valves to their original “on / open” positions