14
If the lid is opened during operation, the cleaning cycle will stop and the following
screen will be displayed:
It is important that you do not turn off the main electricity supply to your
Hygea
2850VM
when it running a cleaning cycle
Your
Hygea 2850VM
recognizes when the main electricity supply fails during a cycle
and prints a report (and/or creates a record on the MC card) when the machine is
next switched back on.
The cycle will be recorded an aborted cycle.
If the cycle fails due to a power failure, the following screen is displayed when the
power is restored:
Important:
the operator must acknowledge this by pressing and holding the “
ENTER
” button to
complete the aborted cycle.
If the MC Card is not present, and “MC Card Only” is selected as the validation mode, then you will
not be able to run another cycle until an MC
card is inserted. A message of “Please insert MC Card”
will appear on the above screen.