47650 Instruction Manual (Rev. 2.1)
4.2
Adjusting the Detection Sensitivity
A small stainless steel screw allows the gap
magnet/keeper to be adjusted thereby enabling the
plate release-sensitivity
to be regulated.
For juvenile (lighter) animals, it might be necessary
to increase sensitivity a little.
Turn the screw CW to increase, CCW to decrease
sensitivity.
Figure 5
“
Detection Sensitivity
”
4.3
Training
It advisable to carry out an acclimation and training session, before starting the motor co-
ordination test.
With the rotor at low speed, say 4-5 R.P.M., each naïve mouse is placed upon its section
in order to familiarize it with the revolving rotor. After 2-3 training runs of 1-2 minutes at
intervals of 2-3 hours, the mice should be accustomed to the rota-rod, and ready for the
test proper.
5
OPERATION
Switch
on
the
Rota-Rod:
the
instrument performs a CPU check,
which might take up to 60 seconds,
during which the display will show:
After switching on the instrument, first of all, get familiar with its controls.
Summary of Contents for 47650