TEMPERATuRE CALIBRATION
A process to set temperature calibration to an ice
point is included to compensate for differences in
temperature sensors, or to reset the unit to proper
operation. Please follow these steps to perform a
temperature set;
1. Connect temperature probe
2. Press “Pressure TEMP” to display probe
temperature (verify that “TEMP” is shown on
the LCD.
3. Place the temperature probe sensing area in a
mixture of crushed ice and water. (Distilled
water will provide the most accurate source)
4. Stir the ice water mixture with the temperature
probe and allow the reading to stabilize
5. Press and hold “Temp SH/SC” button for four
seconds
This will set the ice point to 32°F. The meter will
compensate for changes up to 10°F.