11
IRT807
OFF
ZERO
kHz
Temp
SMOOTH
mV
mA
INSULA
TION
V
V
50V
1000V
INSULATION TESTER
Apo
MIN/MAX
LOCK
Hz
Test Voltage
F C
SELECT
TEST
RANGE
PI/DAR
A-HOLD
0
0.1 M
1 M
10 M
100 M
1 G
2 G
V
DC
M
IRT807
Apo
F
IRT807
OFF
ZERO
kHz
Temp
SMOOTH
mV
mA
INSULA
TION
V
V
50V
1000V
INSULATION TESTER
Apo
MIN/MAX
LOCK
Hz
Test Voltage
F C
SELECT
TEST
RANGE
PI/DAR
A-HOLD
0
0.1 M
1 M
10 M
100 M
1 G
2 G
V
DC
M
IRT807
Apo
mV
WARNING
• Use CATIII rated Test leads or higher.
• Do not attempt to measure more than 400mV DC.
• Keep hands below finger guards when measuring high voltage levels.
Features:
MIN/MAX
Test Voltage
F C
RANGE
LOCK
A-HOLD
SELECT
DC MILLIVOLTS
RANGE
RESOLUTION
ACCURACY
400mV DC
0.1mV
±(0.2% + 3 dgts)
Temperature: -328˚ to 2462˚F (-200˚ to 1350˚C)
Fahrenheit is the default temperature measurement scale. Press the
Range Button
to switch to Celsius scale.
TEMPERATURE
RANGE
RESOLUTION
ACCURACY
1
-328˚ to 999˚F (-200˚ to 999˚C)
0.1˚F (0.1˚C)
±(1.5% +3.6˚F (1.5% +2˚C))
1000˚ to 2642˚F (1000˚ to 1350˚C)
1˚F (1˚C)
±(1.5% +3˚F (1.5% +2˚C))
1
Accuracies apply following 90 minutes setting time after a change in the ambient temperature of the instrument
DC mV: 1mV to 400mV
• Default =
mV
• Press x1 =
Temp
• Default =
mV
• Press x1 =
Temp
mV
mV