Earth-bond Resistance

IRT807

OFF

ZERO

kHz

Temp

SMOOTH

mV

mA

INSULA

TION

V

V

50V

1000V

INSULATION TESTER

Apo

MIN/MAX

LOCK

Hz

Test Voltage

F  C

SELECT

TEST

RANGE

PI/DAR

A-HOLD

0

0.1 M

1 M

10 M

100 M

1 G

2 G

V

DC

   M

IRT807

Apo

•  Default   =  
•  Press x1 =  

ZERO

SELECT

NOTE: 

Measurements can be adversely affected by impedances of additional operating circuits connected in parallel or    

 

            transient currents.

1.  Insert test leads into the Zero/Ω and Com input ports.
2.  Turn the rotary dial selector to the Zero/Ω position.
3.  Short the test leads together. Press the 

SELECT Button

 and wait until ZERO appears on the display. The tester measures the    

  probes resistance, stores the value in memory and subtracts it from tests. The probes resistance is saved even when the tester  

  is powered off. If the probes resistance is >2Ω, the resistance will not be saved.
4.  Connect the test leads to the circuit to be tested. The tester will automatically detect if the circuit is energized.
5.  The dashes (----) will display until the 

TEST Button

 is pressed and a valid resistance value is detected.

6.The High Voltage symbol will display if greater than 2V AC/DC is present. 
  If this is happens, the test is corrupted. Disconnect the test leads and remove power before testing again.
7.  Press and hold the 

TEST Button

 on the Meter or Remote Probe to start the test. The TEST symbol will appear on the display    

  until the 

TEST Button

 is released. The resistance value is displayed until a new test is started or a different function or range  

  is selected.

Features:  

TEST

LOCK

A-HOLD

SELECT

 

 NOTE:

 

When the resistance is higher than the maximum display range the tester displays the > symbol and the maximum  

 

             resistance for the range.

 WARNING

NEVER test resistance on a live circuit. (Overload Protection: AC 2V rms or DC)

RANGE

RESOLUTION

ACCURACY 

1

20.00Ω

0.01Ω

±(1.5% +3 dgts)

200.0Ω

0.1Ω

2000Ω

20.00kΩ

0.01kΩ

1. Accuracies apply from 0 to 100% of range

2. Open Circuit Test Voltage: > 4.0V, < 8V

3. Short Circuit Current: > 200.0 mA

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for IRT807

Page 1: ...STER Apo MIN MAX LOCK Hz Test Voltage F C SELECT TEST RANGE PI DAR A HOLD 0 0 1M 1M 10 M 100 M 1 G 2 G VDC M INSTRUCTION MANUAL ENGLISH IRT807 1000V AC DC Digital Insulation Resistance Tester www Glob...

Page 2: ...328 to 2462 F 200 to 1350 C 11 Resistance 12 Continuity 12 Diode Test 13 Capacitance 13 mAmps AC DC 14 Frequency 14 Earth bond Resistance 15 Insulation Resistance 1000V 500V 250V 100V 50V 16 Measuring...

Page 3: ...emperature 14 to 122 F 10 to 50 C Operating Humidity 75 Operating Altitude 6 526 ft 2000m Pollution Degree 2 Over range OFL is displayed Backlight Yes Refresh Rate 3 sec Dimensions 8 27 x 3 52 x 2 17...

Page 4: ...ment PPE to prevent shock and arc blast injury where hazardous live conductors are exposed Always turn off power to a circuit or assembly under test before cutting unsoldering or breaking the current...

Page 5: ...e test Continuity SMOOTH INSULATION 50V 1000V Insulation Test mV DCmV PI DAR Polarization Index Dielectric Absorption Ratios mA Milliamp AC DC Auto HOLD ZERO ZERO Greater Than HOLD Auto power off acti...

Page 6: ...gory Max Indicator Maximum CAT Rating for input jacks Use CATIII test leads or higher Zero Test Lead Port Fused used for Zero Ohms and mA Common Test Lead Port Volt Insulation Test Lead Port CATIV 600...

Page 7: ...er under protective rubber boot Q Serial Number under kick stand Default Press x1 k Voltage 1000V AC IRT807 OFF ZERO kHz Temp SMOOTH mV mA INSULATION V V 50V 1000V INSULATION TESTER MIN MAX LOCK Hz Te...

Page 8: ...y generated by frequency motor drives and inverters WARNING Use CATIII rated Test leads or higher Do not attempt to measure more than 1000V AC Keep hands below finger guards when measuring high voltag...

Page 9: ...es MIN MAX Test Voltage F C RANGE LOCK A HOLD SELECT DC MILLIVOLTS RANGE RESOLUTION ACCURACY 400mV DC 0 1mV 0 2 3 dgts Temperature 328 to 2462 F 200 to 1350 C Fahrenheit is the default temperature mea...

Page 10: ...PI DAR A HOLD 0 0 1M 1M 10 M 100 M 1 G 2 G VDC M IRT807 Apo Default Press x1 Press x2 Press x3 SELECT IRT807 OFF ZERO kHz Temp SMOOTH mV mA INSULATION V V 50V 1000V INSULATION TESTER Apo MIN MAX LOCK...

Page 11: ...C RANGE CAPACITANCE RANGE RESOLUTION ACCURACY OVERLOAD PROTECTION 10 00nF 0 01nF 2 5 5 dgts 600V 100 0nF 0 1nF 1 000 F 0 001 F 10 00 F 0 01 F 100 00 F 0 1 F 1000 F 1 F 3 5 dgts 10000 F 1 F WARNING Ca...

Page 12: ...AC 45Hz to 1000Hz 40mA 0 01mA 1 5 5 dgts 1 400mA 0 1mA DC 40mA 0 01mA 0 5 5 dgts 400mA 0 1mA 1 1kHz bandwidth Overload Protection 440mA 1000V FAST fuse IRT807 OFF ZERO kHz Temp SMOOTH mV mA INSULATION...

Page 13: ...ect if the circuit is energized 5 The dashes will display until the TEST Button is pressed and a valid resistance value is detected 6 The High Voltage symbol will display if greater than 2V AC DC is p...

Page 14: ...ust be connected in parallel to the circuit or circuit element under test Set Rotary dial to INSULATION and press RANGE button to select desired voltage The display will show until the button is press...

Page 15: ...unction dial to INSULATION position and press RANGE button to select desired voltage Press the PI DAR button to select PI or DAR The display shows until TEST button is pressed and a valid resistance i...

Page 16: ...A 1000V 1R 10kA Replace the battery cover Replace the protective rubber boot Fuse AA battery AA battery AA battery AA battery Fuse AA battery AA battery AA battery AA battery Removed protective rubber...

Page 17: ...ive from such defects the unit will be repaired or replaced at UEi s option This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration...

Reviews: