ANALYZER PROBLEM SOLVING
If any problems are not solved with these solutions, contact us or an authorized repair center .
FAULT SYMPTOM
CAUSES / SOLUTIONS
Oxygen too high
CO
2
too low
Air leaking into probe, tubing, water trap, connectors or internal to analyzer .
Batteries not holding charge
Analyzer not running on mains adapter .
Batteries exhausted .
AC charger not giving correct output .
No fuse
Analyzer does not respond to flue gas
Particle filter blocked .
Probe or tubing blocked .
Pump not working or damaged with contaminants .
Net temperature or Efficiency calculation incorrect .
Ambient temperature set wrong during Automatic Calibration
Flue temperature readings erratic
Temperature plug reversed in socket .
Faulty connection or break in cable or plug .
T flue or ΔT displays (-N/F-)
Probe not connected .
Faulty connection or break in cable or plug .
EFF or X-Air displays (- O
2
2++-)
CO
2
reading is below 2% .
0
2
> 18%
Analyzer just continually beeps
Turn dial back to “MENU” and press “ENTER”
Turn dial back to “EXCH TEST” and press “ENTER”
FREQUENTLY ASKED QUESTIONS
Q:
What is the countdown time on a C165
A:
There are two levels of countdown (aka fresh air purge) on a C165 .
From first switch on if ‘cold’ (more than 41°F from the temperature at which calibrated) = 90 secs .
From first switch on if ‘warm’ (within 41°F of the temperature at which calibrated) = 60 secs
PRODUCT REGISTRATION
Register Your Analyzer
Product registration allows us to contact you regarding potential updates or safety information related to your analyzer . It also speeds up the
recertification process when you send your analyzer in for annual service . Finally users have the option to be notified of related news or new
product information .
Register Online
Visit UEITEST .COM
Click the Product Registration banner on the home page & fill out the form
Upload a scan or photo of your proof of purchase
Enter the Captcha form & send
Register by Mail
Fill out the form at the back of this manual
Mail it to UEi Test Instruments
Find the Serial Number by removing the protective boot from the analyzer
The serial number will be found on the back label