AUTOLOG TIME:

  Can be set from 5 secs to 300 secs.

  

MENU 

  

 

 

 

AUTOLOG TIME 

= 10 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

 

12.52.30 pm 

03/22/15 

      

 

  

NOTE:

  

If the PUMP is switched off then AUTOLOG is inhibited

START AUTOLOG

TIP: 

 

Before starting AUTOLOG you might want to manually re-zero the analyzer so that you have 30 

minutes of logging time before a re-zero is requested.

Switches ON or OFF automatic timed logging

  

MENU 

  

 

 

 

START AUTOLOG? 

YES 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

  

 

 

 

12.52.30 pm 

03/22/15 

      

 

  

To initiate AUTOLOG go back to the main display where the Autolog time will be displayed in the top 
right hand corner of the display, and then press 

. The analyzer will beep and the autolog time will 

disappear for a second or so.

AUTOLOG can be toggled on and off by pressing and holding 

 until the autostore time in the top right 

hand side of the screen appears or disappears.

Summary of Contents for AUTOPLUS 4-2

Page 1: ...YSER ZERO CAL TIME 13 01 17 35PM 18 04 12 ppm ppm CO2 CO O2 HC NO AUTOPLUS 4 2 5 2 1 800 547 5740 Fax 503 643 6322 www ueitest com Email info ueitest com INSTRUCTION MANUAL ENGLISH Automotive Exhaust...

Page 2: ...hutdown Sequence 27 5 PRINTING INFORMATION OPTIONAL EXTRA ONLY 28 29 5 1 Printing a Live Test 28 5 2 Standard Printout 29 6 MAINTENANCE 29 30 6 1 Emptying and Cleaning the In line Water Trap 29 6 2 Ch...

Page 3: ...AL PARTICLE FILTER IN LINE WATERTRAP KEEP CLEAR OF WATER AT ALLTIMES WATER LEVEL LIMIT GAS CONNECTION TO PROBE MINI USB PORT SERIAL ON OFF UP Scrolls up through options MENU Allows access to all menu...

Page 4: ...LAYOUT REAR AUTOplus 3 2 AUTOplus 4 2 AUTOplus 5 2 KANE INTERNATIONAL LTD Swallowfields Welwyn Garden City Herts AL7 1JG Tele 44 0 1707 375550 Fax 44 0 1707 393277 Made in the UK Serial No BUILD STATU...

Page 5: ...ssary for on road rolling road vehicle testing when the temperature of the emission gases increase Please request further details from your distributor or contact UEi directly Probe Inlet Water Trap C...

Page 6: ...nstallation category II Pollution degree 2 Indoor use only Altitude to 2000m Ambient temperature 0 C 40 C Maximum relative humidity 80 Mains supply fluctuations not to exceed 10 of the nominal voltage...

Page 7: ...air the water trap is correctly fitted and the analyzer upright DO NOT RUN Analyzer WITHOUT WATER TRAP FITTED Switch ON the analyzer by pressing 4 2 AUTOMATIC ZERO CALIBRATION During this sequence th...

Page 8: ...g And then the initializing screen is shown FRESH AIR PURGE TIME 90 15 13 27 03 22 15 The time will count down in seconds to zero and is pre set by the analyzer During countdown the analyzer will init...

Page 9: ...ed and then after a further minute beep continuously The bottom line of the display shows time date and battery level When a zero is required the top left hand corner of the display will show ZERO REQ...

Page 10: ...en not sampling to preserve battery life PRINT Also acts as SEND key for Bluetooth and serial communications when the appropriate printer option is selected When active PRINTING appears in the top lef...

Page 11: ...om the MENU function Press and hold for 2 seconds to toggle AUTOLOG on and off NOTE AUTOLOG is disabled when the pump is switched off 4 4 MENU There are 5 main menu categories each with their own sub...

Page 12: ...MENU SETUP UNITS SCREEN REPORTS SERVICE MANUAL ZERO 12 52 30 pm 03 22 15 Use to scroll the cursor through the selection and the press to choose to exit this screen press MENU again SETUP MENU LANGUAG...

Page 13: ...GE Use to change Use to accept MENU LANGUAGE English French German Dutch 12 52 30 pm 03 22 15 PUMP Pump speed can be set from 1 slowest to 4 fastest Use to change Use to accept MENU PUMP 4 12 52 30 pm...

Page 14: ...ange Use to accept MENU PRINTER KANE IRP 2 KANE IRP USB SERIAL BLUETOOTH 12 52 30 pm 03 22 15 SET TIME Use to change Use to accept and move the cursor MENU SET TIME HH MM SS 12 52 30 am 12 52 30 pm 03...

Page 15: ...SET DATE DD MM YY 26 04 12 12 52 30 pm 03 22 15 NOTE If FUNCTION LOCKED appears delete all reports see p20 22 Date format can also be changed before the date by using the same method PASSKEY This nee...

Page 16: ...se to change Use to accept MENU FUEL VEHICLE EFF PEF MEF BACK 12 52 30 pm 03 22 15 FUEL NOTE when diesel is selected HC and Lambda are not displayed MENU FUEL gasoline petrol LPG CNG DIESEL 12 52 30 p...

Page 17: ...allows a vehicle registration number to be entered Use to change Use to accept MENU VEHICLE VE05DPZ 12 52 30 pm 03 22 15 EFF Set Lambda or AFR Use to change Use to accept MENU EFF LAMBDA AFR 12 52 30...

Page 18: ...ays value only MENU PEF 0 545 12 52 30 pm 03 22 15 Example 1000ppm of HC hexane PEF 1835ppm propane MEF Methane Equivalency Factor displays value only MENU MEF 0 0545 12 52 30 pm 03 22 15 Example 1000...

Page 19: ...19 SCREEN MENU CONTRAST BACKLIGHT LINES BACK 12 52 30 pm 03 22 15 Use to change Use to accept CONTRAST set contrast from 1 to 15 MENU CONTRAST 12 12 52 30 pm 03 22 15...

Page 20: ...m 03 22 15 LINES Allows the main display to be configured by the user MENU LINE 1 CO2 LINE 2 LINE 3 LINE 4 LINE 5 LINE 6 EXIT 12 52 30 pm 03 22 15 NOTE If FUNCTION LOCKED appears delete all reports se...

Page 21: ...LOG HEADER 1 HEADER 2 BACK 12 52 30 pm 03 22 15 VIEW REPORTS MENU LOG 13 Date 07 10 11 Time 03 35 VEHICLE VE05DPZ FUEL DIESEL CO2 14 4 CO 0 05 O2 1 1 HC NO 150 ppm 12 52 30 pm 03 22 15 Reports can be...

Page 22: ...22 DELETE ALL REPORTS MENU DELETE ALL NO YES 12 52 30 pm 03 22 15 MENU CONFIRM DELETE NO YES 12 52 30 pm 03 22 15...

Page 23: ...efore a re zero is requested Switches ON or OFF automatic timed logging MENU START AUTOLOG YES 12 52 30 pm 03 22 15 To initiate AUTOLOG go back to the main display where the Autolog time will be displ...

Page 24: ...1 HEADER 2 allows inputting of company details to appear on two lines of the printout MENU HEADER 1 Company Name 12 52 30 pm 03 22 15 MENU HEADER 2 Phone number 12 52 30 pm 03 22 15 Use to change Use...

Page 25: ...CODE entry is reserved for approved service agents CAL DATE is the countdown for days remaining until annual calibration is due MANUAL ZERO This allows the user to manually re zero the analyzer REMOVE...

Page 26: ...attached to the exhaust pipe 4 6 TURNING THE PUMP ON OFF The analyzer is fitted with a pump to draw gas from the vehicle exhaust To conserve battery power switch off the pump when you are not taking a...

Page 27: ...become blocked If this filter is allowed to become dirty then damage may occur inside the analyzer Check the following for water Probe line water trap and particle filter When the blockage is cleared...

Page 28: ...a cable to transmit the data but uses an infrared IR link similar to a TV remote control The IR emitter is positioned on the top of the analyzer and the bottom of the printer Ensure they are pointing...

Page 29: ...and printed TIP Make a note of the location number for your particular test as it may be useful when printing TIP Stored and displayed with the data are actual time and date of the test 6 MAINTENANCE...

Page 30: ...with a damped cloth rinsed in water Leave the analyzer in a warm place until all surface moisture has evaporated and the analyzer is completely dry Under no circumstances should you use solvent based...

Page 31: ...be faulty Analyzer does not respond to exhaust gas Particle filter blocked Probe or tubing blocked Pump not working damaged with contaminants or switched off Analyzer automatically switches off in ope...

Page 32: ...DISPLAY 9 PRODUCT SPECIFICATION FOR HANDHELD AUTOPLUS 4 2 AUTOPLUS 5 2 Parameter Resolution Accuracy Range Carbon Monoxide Infrared 0 01 5 of reading 1 0 06 volume 1 0 10 Over range 20 Oxygen fuel cel...

Page 33: ...t standard temperature and pressure STP NOTE The analyzer is not for use with a Dynamometer unless the high temperature probe is fitted APPENDICES A MAIN DISPLAY PARAMETERS The parameters and their me...

Page 34: ...alculation for the AFR is Lambda multiplied by 14 7 for Petrol and 15 6 for LPG typically When sampling fresh air this indication will show NO Nitric oxide reading in ppm parts per millions of the exh...

Page 35: ...is 6 10 4 HCV Atomic ratio hydrogen to carbon in the fuel Nominal value is 1 7261 OCV Atomic ratio oxygen to carbon in the fuel Nominal value is 0 0176 C ELECTROMAGNETIC COMPATIBILITY CE STATEMENT Th...

Page 36: ...a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any cl...

Reviews: