31
If you suspect the internal filter is blocked perform the following:
• Remove the probe connection from the water trap.
• Empty and clean the water trap with a dry cloth.
• Fit a new external particle filter.
• Run the analyzer in fresh air (pump ON) for at least one hour.
If the problem does not clear contact a service agent
Fault symptom
Causes
• Oxygen too high
• Air leaking into probe, tubing, water trap,
connectors or internal to analyzer.
• Oxygen cell needs replacing.
• Oxygen Error (FAULT)
• Infrared gas Error (FAULT)
• Zero calibration fault
• Analyzer has been stored in a cold
environment and is not at normal working
temperature.
• Oxygen cell or infrared bench needs replacing.
• Analyzer not holding charge
• Analyzer not charging
• Battery exhausted.
• Charger not giving correct output.
• Charger could be faulty.
• Analyzer does not respond to exhaust gas
• Particle filter blocked.
• Probe or tubing blocked.
• Pump not working, damaged with
contaminants or switched off.
• Analyzer automatically switches off in
operation
• Battery below alarm level.
• Ambient temperature above 50°C.
• Battery quickly discharging and is faulty.
8. ZERO CHECKS AND RE-CALIBRATION
During normal operation of the analyzer the following checks may be requested as required:
• Zero setting of all sensors (can also be selected manually)
8.1 ZERO SETTING
The zero setting function sets the working sensors to zero using fresh air. This function is activated as
follows:
• Following analyzer switch ON.
• On a timed basis. Following the analyzer first being turned ON a zero will requested automatically at
15 and then 30 minute intervals. Subsequent requests will be every 30 minutes.
• On request by the user from the
MENU
The zero sequence is as follows, ensure the on screen commands are followed or the analyzer may not zero
correctly.