Technical information
EN-59
english
E
Technical information
Serial number:
The unit bears a serial number SN which provides clear identification.
Power supply:
Battery type: 4 x 1.5 V AA alkali manganese (LR 6) or lithium (FR 6) round cells, service life:
more than 800 measurements in 2 years
Optionally stabilised type A1 mains adapter, output 6 VDC, minimum 600 mA
Cuff:
Cuff VM5 (for arm circumference 22-32cm) or optionally
cuff VL5 (for arm circumference 32-42 cm)
Operating conditions:
Ambient temperature 10 to 40 °C, relative humidity up to 85 %, non-condensing
Storage and transport
conditions:
Ambient temperature -20 to +50 °C, relative humidity up to 85 %, non-condensing
Air deflation valve:
Electronically controlled
Automatic switch-off
approx. 3 minutes after end of measurement
IP code:
IPX0 not protected
Technical data
3.
Original spare parts and accessories:
4.
Subject to technical modifications.
Cuff 22-32 cm type VM5
Part. no. 2505001
PZN 9101599
•
Mains adapter type A1
Part. no. 2501020
PZN 0179482
•
Cuff 32-42 cm type VL5
Part. no. 2505002
PZN 9101607
•
Unit type:
Digital automatic unit with electric pump for measuring blood pressure on upper arm
Dimensions:
L = 100 mm x W = 152 mm x H = 61 mm
Weight:
295 g not including batteries
Display:
LCD display (liquid crystal display) for measured values and check displays
Memory:
60 measured values (stored automatically) and average value (A)
Measurement
procedure:
Oscillometric determination of systole, diastole and pulse
Reference procedure
of clinical testing:
Auscultatory measurement
Inflating pressure:
Initially up to approx. 180 mmHg, then in stages of 30 mmHg
Pressure display
range:
0-300 mmHg
Measurement range
Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse: 40-160 pulses/min
Error limits:
Blood pressure measurement: corresponds to EN 1060 Part 3
Pressure measurement: ± 3 mmHg
Pulse measurement: ± 5%