Failure encountered
Possible cause
Corrective action
Measured values too high
Too little rest before the measurement.
Repeat measurement after a break of
approx. 3-5 mins.
- Do not move arm
- Do not talk
Unusual measured values
Movement or talking during measurement,
resting time not observed, feet possibly
crossed, smoking or drinking coffee.
Check conditions and repeat measurement.
Observe instructions for use P. EN-44.
Display 0
0
Start/Stop button accidentally pressed while
battery was being changed.
Switch unit off and then on again with
Start/Stop button; if necessary.
Measurement was
interrupted and cuff first
deflated and then inflated
Unit detects a fault or that diastolic value
is too low.
Unit shows no fault. Repeat measurement.
Movement during measurement
If necessary, interrupt measurement, meas-
ure again after a break of 5 minutes.
No display after unit is
switched on
Batteries inserted incorrectly.
Check position of batteries.
Batteries flat.
Change batteries.
Battery contacts dirty.
Clean battery contacts.
No display after Memory
button is pressed
No measured values stored.
Take a new measurement.
Measurement interrupted
Batteries flat
Change batteries.
Failure encountered
Possible cause
Corrective action
Measured values extremely
high or low
Incorrect measurement position.
Repeat measurement, ensuring correct body
posture. Do not move your arm and do not
talk.
Date/time not shown in
display
- Function accidentally deactivated
- Function was not reactivated after battery
replacement
Activate date and time as described on P.
EN-46.
Customer service
2.
The unit may only be repaired by the manufacturer or by an expressly authorised organisation. Please contact: