33
FIELD ENTRY LIST FOR STEAM:
Test Results: OK/LEK/BLW/NIS/PLG/SIZ/OTH
Man: SAR/ARM/BES/NIC/
Model: MDLO to MDL9
App: SM/AH/RAD/HE/TR/
Type: IB/TD/TH/FT/
PIPE DIA:
1/16, 1/8, ¼, 3/8, 1/2, ¾, 1.00, 1.25, 1.50, 2.00, 2.50, 3.00, 3.50, 4 up to 36 in.
ORIFICE SIZE 1/32, 1/16, 3/32, 1/8, 5/32, 3/16, 7/32, ¼, 9/32, 10/32, 11/32, 3/8, 13/32
F. Leaks
1. To view
Leak Info
, click the
Sensitivity Control Dial
to move the cursor to the bottom of the
Display Panel
until
“Press
ENTER For”
is blinking. Spin to
Leak Info
and press Enter.
2. To enter a test condition:
3. Spin to
Test Results
: the condition default is
LEK
and will blink slowly. If it is necessary to change:
4. Click with the
Sensitivity Control Dial
and the condition (
LEK
) will blink rapidly.
5. Spin to select a condition indication. The choices are;
LEK, CHK, OK
6. To Set, Click the
Sensitivity Control Dial
(the condition will blink slowly)
7. Spin to
Pressure
8. Click on
Pressure
to enter and spin to desired pressure (
PSI
or
BAR
)
9. Click to Set
10. Spin to
Application
(codes as
APP0-APP9
)
11. Click to Enter and Spin to desired
Application
12. Click to Set
13. Spin to
Distance
14. Click to Enter
15. Spin to select
Distance
16. Click to Set
10. Press
Enter
to store the information and Spin and Click to exit.
If
Parameter Info
has been selected in the
Storage Display
function, the identical data
and data entry as described above will be shown.
Additional information such as
Date
,
Time
,
Module Used
and a command for
Data Storage
will be shown in the
MORE
function.
FIELD ENTRY LIST FOR LEAKS:
Test Results: LEK/CHK/OK
Press:0000 PSI (BAR)
App: App0-9
Distance: 5 in,6 in, up to 15 in, then 2 ft to 30 ft (38 CM , 1M, 1.5M 2M, 2.5M up to 10M
NOTE: All Field Entry Lists shown above are the factory lists that can be changed/customized in the Ultratrend
Software. After the changes have been completed, they may be uploaded to the Ultraprobe 10000.
Press ENTER LEAK
To store
009
Test Results: LEK
Prss: 0000 PSI
App: App 0
Distance: 5 in
60 dB
40 kHz
S=69
RT
Press ENTER
for Leak Info