24
Text Editor = ON
Alarm = Disabled
Module Type Select = SCM
Display Response = Medium
Sensitivity Mode = Manual
Headphone Volume = 99%
Turnoff Time = 5 Seconds
Shutoff Mode = Timed
The user defined Default Settings can be:
User Sensitivity Default (see Setup Menu 13)
User Frequency (see Setup Menu 14)
Frequency Adjust (see Setup Menu 15)
To select Default Settings:
1.
Follow the steps to Enter the
Setup Mode
as described earlier.
2.
Spin to
Menu12
:
Default Settings
3.
Click to Enter
4.
Spin to
“YES”
or
“NO”
5.
Click to set
6.
Press the
Yellow ENTER button
to exit.
Menu 13: Sensitivity Default
The user can select a default sensitivity level ranging from 0-70
1.
Follow the steps to Enter the
Setup Mode
as described earlier.
2.
Spin to
Menu 13
:
Sensitivity Default
3.
Click to Enter
4.
Spin to
selected sensitivity level
5.
Click to set
6.
Press the
Yellow ENTER button
to exit.
Menu 14: User Frequency Default
This is the frequency to which the instrument will automatically default when turned on.
1.
Follow the steps to Enter the
Setup Mode
as described earlier.
2.
Spin to
Menu 14
:
Frequency Default
3.
Click to Enter
4.
Spin to
desired frequency
5.
Click to set
6.
Press the
Yellow ENTER button
to exit.
Menu 15: Frequency Adjust
Select
YES
to manually change frequency during operations. Select
NO
to fix a selected (default) frequency to be used during all
inspections.
1.
Follow the steps to Enter the
Setup Mode
as described earlier.
2.
Spin to
Menu 15
:
Frequency Adjust
3.
Click to Enter
Menu
12
Select
Default Settings
Press ENTER to Exit
Default Settings
Confirm = (YES)
Default Settings
Confirm = (NO)