3
|
When arc calibration is completed, the following screen is displayed.
4
|
Press to stop arc before calibration is completed if necessary.
6.5.2 ARC TEST
Swift KF4 measures by ARC TEST menu if arc power is adequate. When arc power is too weak or too strong it is
automatically adjusted by executing ARC CALIBRATION menu.
Only SM fi ber should be used for arc calibration.
1
|
Prep and insert SM fi ber into the splicer using clamps or proper fi ber holders.
2
|
Press as follows.
35
36
6.5
|
CALIBRATION
To display splice mode, press and select the “CALIBRATION” menu with the button. The calibration
menu is equipped with various functions, such as arc amount calibration, motor operation test, etc.
6.5.1 ARC CALIBRATION
The KF4 continuously checks for a change in temperature and air pressure through each sensor. Based on such
data, arc amount is automatically adjusted. However, a change in arc amount due to abrasion of the electrodes
or the fi ber splice, is not automatically adjusted. The central axis of the arc can also be moved toward the left or
right. In this case, arc calibration is required.
When executing arc calibration, arc voltage is automatically changed to a proper value. This value is calculated
internally. The arc voltage cannot arbitrarily be changed.
Only SM fi ber should be used for arc calibration.
1
|
Prep and insert SM fi ber into the splicer using clamps or proper fi ber holders.
2
|
Press as follows.
SWIFT KF4-EN, Rev.C 9
3.
Product Outline
3.1.
Function buttons
BUTTON
DESCRIPTION
Press and hold about 1 second to turn the power on/off.
Move the cursor to the left.
Move fiber to manual mode and adjust the camera’s focus.
Stripping popup menu should load.
Move the cursor to the right.
Move fiber to manual mode and adjust the camera’s focus.
Move the cursor upward.
Move each motor to manual mode.
Splice popup menu should load.
Move the cursor downward.
Move each motor to manual mode.
Heater popup menu should load.
Initialize the splice function.
Return to the menu screen.
Complete a selection.
Follow the next step on the menu screen.
Splice execution.
Return to the initial screen and initialize splice function.
Turn on the heater.
When it is on, the lamp on the left is in the red.
Press once more when it is on, and the heater is turned off.
SWIFT KF4-EN, Rev.C 9
3.
Product Outline
3.1.
Function buttons
BUTTON
DESCRIPTION
Press and hold about 1 second to turn the power on/off.
Move the cursor to the left.
Move fiber to manual mode and adjust the camera’s focus.
Stripping popup menu should load.
Move the cursor to the right.
Move fiber to manual mode and adjust the camera’s focus.
Move the cursor upward.
Move each motor to manual mode.
Splice popup menu should load.
Move the cursor downward.
Move each motor to manual mode.
Heater popup menu should load.
Initialize the splice function.
Return to the menu screen.
Complete a selection.
Follow the next step on the menu screen.
Splice execution.
Return to the initial screen and initialize splice function.
Turn on the heater.
When it is on, the lamp on the left is in the red.
Press once more when it is on, and the heater is turned off.
SWIFT KF4-EN, Rev.C 48
6.5.
CALIBRATION
To display splice mode, press
and select the “CALIBRATION” menu with the
button. The calibration menu is equipped with various functions, such as arc amount calibration,
motor operation test, etc.
6.5.1 Arc calibration
The KF4 continuously checks for a change in temperature and air pressure through each sensor.
Based on such data, arc amount is automatically adjusted. A change in arc amount due to
abrasion of the electrodes or the fiber splice, however, is not automatically adjusted. The central
axis of the arc can also be moved toward the left or right. In this case, arc calibration is required.
- When executing arc calibration, arc voltage is automatically changed to a proper value. This
value is calculated internally; the arc voltage cannot arbitrarily be changed.
- Only SM fiber should be used for arc calibration.
SWIFT KF4-EN, Rev.C 48
6.5.
CALIBRATION
To display splice mode, press
and select the “CALIBRATION” menu with the
button. The calibration menu is equipped with various functions, such as arc amount calibration,
motor operation test, etc.
6.5.1 Arc calibration
The KF4 continuously checks for a change in temperature and air pressure through each sensor.
Based on such data, arc amount is automatically adjusted. A change in arc amount due to
abrasion of the electrodes or the fiber splice, however, is not automatically adjusted. The central
axis of the arc can also be moved toward the left or right. In this case, arc calibration is required.
- When executing arc calibration, arc voltage is automatically changed to a proper value. This
value is calculated internally; the arc voltage cannot arbitrarily be changed.
- Only SM fiber should be used for arc calibration.
SWIFT KF4-EN, Rev.C 9
3.
Product Outline
3.1.
Function buttons
BUTTON
DESCRIPTION
Press and hold about 1 second to turn the power on/off.
Move the cursor to the left.
Move fiber to manual mode and adjust the camera’s focus.
Stripping popup menu should load.
Move the cursor to the right.
Move fiber to manual mode and adjust the camera’s focus.
Move the cursor upward.
Move each motor to manual mode.
Splice popup menu should load.
Move the cursor downward.
Move each motor to manual mode.
Heater popup menu should load.
Initialize the splice function.
Return to the menu screen.
Complete a selection.
Follow the next step on the menu screen.
Splice execution.
Return to the initial screen and initialize splice function.
Turn on the heater.
When it is on, the lamp on the left is in the red.
Press once more when it is on, and the heater is turned off.
SWIFT KF4-EN, Rev.C 49
i.
Prep and insert SM fiber into the splicer using clamps or proper fiber holders.
ii.
Press
as follows.
iii.
When arc calibration completes, the following screen is displayed.
iv.
Press
to stop arc before calibration is completed if necessary.
SWIFT KF4-EN, Rev.C 49
i.
Prep and insert SM fiber into the splicer using clamps or proper fiber holders.
ii.
Press
as follows.
iii.
When arc calibration completes, the following screen is displayed.
iv.
Press
to stop arc before calibration is completed if necessary.
SWIFT KF4-EN, Rev.C 9
3.
Product Outline
3.1.
Function buttons
BUTTON
DESCRIPTION
Press and hold about 1 second to turn the power on/off.
Move the cursor to the left.
Move fiber to manual mode and adjust the camera’s focus.
Stripping popup menu should load.
Move the cursor to the right.
Move fiber to manual mode and adjust the camera’s focus.
Move the cursor upward.
Move each motor to manual mode.
Splice popup menu should load.
Move the cursor downward.
Move each motor to manual mode.
Heater popup menu should load.
Initialize the splice function.
Return to the menu screen.
Complete a selection.
Follow the next step on the menu screen.
Splice execution.
Return to the initial screen and initialize splice function.
Turn on the heater.
When it is on, the lamp on the left is in the red.
Press once more when it is on, and the heater is turned off.
SWIFT KF4-EN, Rev.C 50
6.5.2 Arc Test
Swift KF4 measures by ARC TEST menu if arc power is proper. When arc power is too weak or too
strong it is automatically adjusted by executing ARC CALIBRATION menu.
l
Only SM fiber should be used for arc calibration.
i.
Prep and insert SM fiber into the splicer using clamps or proper fiber holders.
ii.
Press
as follows.
iii.
When arc test completes, the following screen is displayed.
!
CAUTION
SWIFT KF4-EN, Rev.C 9
3.
Product Outline
3.1.
Function buttons
BUTTON
DESCRIPTION
Press and hold about 1 second to turn the power on/off.
Move the cursor to the left.
Move fiber to manual mode and adjust the camera’s focus.
Stripping popup menu should load.
Move the cursor to the right.
Move fiber to manual mode and adjust the camera’s focus.
Move the cursor upward.
Move each motor to manual mode.
Splice popup menu should load.
Move the cursor downward.
Move each motor to manual mode.
Heater popup menu should load.
Initialize the splice function.
Return to the menu screen.
Complete a selection.
Follow the next step on the menu screen.
Splice execution.
Return to the initial screen and initialize splice function.
Turn on the heater.
When it is on, the lamp on the left is in the red.
Press once more when it is on, and the heater is turned off.
SWIFT KF4-EN, Rev.C 50
6.5.2 Arc Test
Swift KF4 measures by ARC TEST menu if arc power is proper. When arc power is too weak or too
strong it is automatically adjusted by executing ARC CALIBRATION menu.
l
Only SM fiber should be used for arc calibration.
i.
Prep and insert SM fiber into the splicer using clamps or proper fiber holders.
ii.
Press
as follows.
iii.
When arc test completes, the following screen is displayed.