18
2. Fast drying mode
This mode is suitable for samples with high moisture content. The
heating temperature can be set, and the initial temperature will exceed
the set value, and will later drop to the set temperature and remain until
the test ends automatically.
3. Soft drying mode
This mode is suitable for samples with low moisture content. The time
required for the heater to rise to the set temperature is longer, and the
temperature rises evenly and slowly. The time during the temperature
rise phase can be set manually.
5.1.3 Ending Mode setting
After drying mode is set, the icons for ending mode and the currently set
drying mode flashes. Press the
“
UP
”
or
“
DOWN
”
to select required ending
mode and confirm by pressing
“
CONFIRM
”
.