V. Using the Balance
Level adjustment
Once the balance is placed at its location, check the bubble level; if bubble is not
well-centered, turn the adjustable feet so the bubble moves to the center and the
balance is properly leveled.
Start
1.
Connect the balance to the power supply and press ON key on the panel.
Wait 30 minutes after powering on for best performance.
2.
After warming up, calibrate the balance.
3.
After calibration, the balance will enter weighing mode
4.
To turn the balance off, press the OFF key on the panel and the display will shut
off. If the balance will be unused for an extended period of time, disconnect it
from the power supply.
Calibration
1.
Press CAL key; the value of the necessary weight will flash on display
2.
Put the calibration weight on the pan. Wait until the value of the calibration
weight stops flashing and is fixed on the display (e.g. 100.0000 g, 200.0000g,
etc., depending on the model)
3.
Remove the weight from the pan
4. The balance will enter weighing mode and will display 0.000g or 0.0000g
Note: Each balance includes the necessary weight to complete external calibra-
tion
Tare function
1.
Put the receptacle on the weighing pan; its weight will be displayed
2.
Press TARE key and the reading on the display will automatically be set to
0.000g or 0.0000g; tare is complete
3.
If the receptacle is removed from the pan, the corresponding weight, with
negative value will be displayed.
4.
Press the TARE key again and the value on display will be set to zero again
Weighing unit selection
Press UNIT key to select the desired weighing unit from g, ct, oz, and lb
Piece counting function
1.
Press the PCS button
2.
Press the CAL button repeatedly to cycle through different counts. This function
can be used with counts of 10, 20, 50, 100, 150, 200, 250, or 500 pieces.
3.
Put the corresponding number of pieces on the pan, press the PCS button to
confirm.
4.
Remove the pieces from the pan and the balance is ready to be used in piece
counting mode.
5. To return to weighing mode, simply press the PCS button again.