Single-point calibration follow the sequence
press the button“TARE”, the display shows “0.0000”.
press“CAL”b
u
tton
,
balance
enter
s
into
calibration
procedure,
the
display
show
s
“CAL--0”
,
internal
calibration
balance
will
make
the
calibration
automatically
the
display
shows
from“CAL-0”to
“CAL-200”and
flash
es,
put
on
the
200g
weight
for
FA2204X
model
display shows“200.0000”,remove the weight and finish the calibration.
internal calibration weight value setting
(internal calibration balance)
cho
ose
the
calibration“CAL-2”(follow
3.2.2)
press
“CAL”
b
u
tton,
the
display
shows“CAL-DN”
put
on
the
standard
calibration
weight
accordingly
the
display
shows
“CAL-UP”
,
remove
the
weight.
wait
for
the
calibration
finished
,
display
shows“0.0000”
3. CALIBRATION