PAGE 9 OF 54
0521 IH-5276
DYNAMIC WEIGHING
Use this application to weigh an unstable load, such as
a moving animal. Two different start/reset modes can
be selected:
Manual
(start and stop via key press) and
Automatic
(start and stop automatically).
DYNAMIC WEIGHING – MANUAL (DEFAULT)
1. In the upper left portion of the home screen, select
Dynamic Weighing. (See Figure 26)
2. Place objects on the pan and press the
START
button.
3. The balance begins a countdown (averaging
process). During the countdown, the information line
displays the time remaining. If necessary, press
STOP
to quit.
4. When the countdown ends, the result line is
displayed and held. Press
CLEAR
to clear the held
weight and return to the initial screen.
NOTE: Clear the pan before beginning a new
dynamic weighing cycle.
ITEM SETTINGS
To view or adjust the current settings, touch the
ITEM SETTINGS
button. The Settings screen appears.
(See Figure 27)
Averaging Time:
Set the averaging time to a value
between 1 and 99 seconds. Default is 5 seconds.
Automatic Mode:
When set to ON, the cycle begins
when an object is placed on the pan, and the held
value is automatically reset when the object is removed
from the pan. (See Figure 28)
Sample Name:
Assign a name to the sample.
Print Settings:
Change printing settings.
NOTE: The SAVE TO USB button will only appear
after a USB flash drive has been connected to
the balance. See page 6 for more information.
DENSITY DETERMINATION
Use this application to determine an object's density.
(See Figure 29)
Four types of density determination can be made:
1. Solids more dense than water
2. Solid less dense than water
3. Liquid density
4. Porous material (impregnated with oil)
OPERATION CONTINUED
Figure 26
0.00
Dynamic Weighing
Menu
Start
Save to
USB
Item
Settings
0.00 g
0.00 g
Gross:
Tare:
5 s
Off
Averaging Time:
Automatic Mode:
>0<
g
Max 4200 g
d=0.01 g
The DYNAMIC
WEIGHING
home screen
Main Display Line
Application
Buttons
Reference Fields
Place sample on the pan. Press Start to begin averaging.
Dynamic Weighing
Print Settings
Sample Name
Automatic Mode
Averaging Time
Exit
Back
Figure 27
Figure 28
0.00
Dynamic Weighing
Menu
Save to
USB
Item
Settings
0.00 g
0.00 g
Gross:
Tare:
5 s
On
Averaging Time:
Automatic Mode:
>0<
g
Max 4200 g
d=0.01 g
Place sample on pan to begin averaging.
Figure 29
0.0000
Density Determination
Menu
Start
Item
Settings
0.00 g
0.00 g
Air Weight:
Liquid Weight:
20.0º C
Off
Water Temp:
Porous Material:
g/cm3
Max 4200 g
d=0.01 g
The DENSITY
DETERMINATION
home screen
Main Display Line
Function Buttons
Reference Fields