- 8 - 

4-3. Rear Panel Control 

17 

Power socket 

 

18 

Fuse holder 

 

19 

AC input selector 

The power transformer is designed to permit operation in 110V (115V/120V) 

or 220V (230V/240V),50/60Hz line voltage. To convert from one line voltage 

to another is done by change AC input selector as shown in section 6-2. 

20 

Cooling fan 

 

 

 

5. OPERATION INSTRUCTIONS 

5-1. Precaution 

1) AC input 

  AC input must be within the range of line voltage ±10% 50/60Hz. 

 

WARNING.  To avoid electrical shock, the power cord protective grounding conductor must be 

connected to ground. 

 

2) Installation 

Allow enough space around the power supply for ventilation. 

 

WARNING. To avoid damaging the power supply, do not use it in a place where ambient 

temperature exceeds 40

 

WARNING. Voltages more than 60V DC are a lethal shock hazard to the user. Be careful when 

connecting power supplies to achieve voltages higher than 60V DC total or 60V DC between any 

connection and earth ground. 

 

 

CAUTION. 

1 The instrument must be operated under rated main supply. If it is meant to work for a long time, 

it is suggested to use 60-70% instead of full load so as to avoid rapid aging.   

2. Avoid frequent short-circuit operations. 

3. In case of the instrument with full load, do not turn on the instrument. Before adding load, adjust 

the voltage knob to the minimum value, then turn on the instrument. Next turn the 

voltage/current adjustment knob to set targeted values. 

 

5-2. Setting Current Limit 

1) Determine the maximum safe current for the device to be powered. 

2) Temporarily short the positive “+” negative “-” terminals of the power supply together with a test lead. 

3) Rotate the COARSE VOLTAGE control away from zero sufficiently for the CC indicator to light. 

4) Adjust the CURRENT control for the targeted current limit. Read the current value on the ammeter. 

5) The current limit (over load protection) has now been preset. Do not change the CURRENT control setting 

after this step. 

6) Remove the short between the “+” and “-” terminals and hook up for constant voltage operation. 

 

Summary of Contents for TP Series

Page 1: ...Single Channel DC Power Supply TP Series Operation Manual V1 0 BENCHTOP INSTRUMENT ...

Page 2: ...CATORS 4 4 1 Panel Illustration 4 4 2 Front Panel Control 7 4 3 Rear Panel Control 8 5 OPERATION INSTRUCTIONS 8 5 1 Precaution 8 5 2 Setting Current Limit 8 5 3 Setting Constant Voltage 9 5 4 C C and C V Mode 9 6 MAINTENANCE 9 6 1 Fuse Replacement 9 6 2 Line Voltage Conversion 10 6 3 Cleaning 10 ...

Page 3: ...to meet the requirements of modern circuit design The unit can be used as either constant voltage or current source It has output ON OFF fixed DC 5V 1A output optional digital LED display and overload protection circuits 2 PRODUCTION MODELS No Model Output Voltage Output Current Output ON OFF Fixed 5V 1A V A Preset Weight kg 1 TP 1303EC 0 30V 0 3A 4 5 2 TP 1305EC 0 30V 0 5A 5 3 TP 1603EC 0 60V 0 3...

Page 4: ... Line regulation 0 2 3mA Load regulation 0 2 3mA 0 2 5mA 0 2 3mA Fixed output Only for TP 1303 1305 1603 Output voltage 5V 1 Output current 1A Ripple noise 3mVrms 5Hz 1MHz Display Voltmeter 3 digits LED display Ammeter 3 digits LED display Resolution 100mV 10mA Accuracy 1 reading 2 digits General Operating environment 0 40 80 RH Storage environment 10 70 70 RH Power source AC110V 220V 10 50 60Hz A...

Page 5: ...oise 3mArms I 3A 10mArms I 3A Display Voltmeter 3 digits LED display Ammeter 3 digits LED display Resolution 100mV 10mA Accuracy Real voltage and current output 1 reading 2 digits Preset voltage and current 1 reading 8 digits General Operating environment 0 40 80 RH Storage environment 10 70 70 RH Power source AC110V 220V 10 50 60Hz Accessories Power cord x1 Operation manual x1 Test lead x1 Dimens...

Page 6: ... 4 4 PANEL CONTROLS AND INDICATORS 4 1 Panel Illustration a TP 1303EC 1305EC 1603EC Front panel Fig 4 1 b TP 1303C 1305C 1603C Front panel Fig 4 2 ...

Page 7: ... 5 c TP 1303 1305 1603 Front panel Fig 4 3 d TP 3010 6005 3020 6010 Front panel Fig 4 4 ...

Page 8: ... 6 e 30V 3A 30V 5A 60V 3A Rear panel Fig 4 5 f TP 3010 6005 Rear panel Fig 4 6 ...

Page 9: ...or Lights when the power supply is in constant current operation 5 C V indicator Lights when the power supply is in constant voltage operation 6 Voltmeter Indicates the output voltage 7 Ammeter Indicates the output current 8 Voltage Coarse For coarse adjustment of the output voltage 9 Voltage Fine For fine adjustment of the output voltage 10 Current Coarse For coarse adjustment of the output curre...

Page 10: ...een any connection and earth ground CAUTION 1 The instrument must be operated under rated main supply If it is meant to work for a long time it is suggested to use 60 70 instead of full load so as to avoid rapid aging 2 Avoid frequent short circuit operations 3 In case of the instrument with full load do not turn on the instrument Before adding load adjust the voltage knob to the minimum value the...

Page 11: ...it automatically changed into constant voltage mode Should check the load or increase the preset voltage value 6 MAINTENANCE WARNING The following instructions are for use by qualified personnel only To avoid electrical shock do not perform any servicing other than contained in the operating instruction unless you are qualified to do so 6 1 Fuse Replacement If the fuse blows the CV or CC indicator...

Page 12: ...he AC line that it may be less than 207V 105V 6 3 Cleaning To clean the power supply use a soft cloth dampened in a solution of mild detergent and water Do not spray cleaner directly onto the instrument since it may leak into the cabinet and cause damage Do not use chemicals containing benzene benzene toluene xylene acetone or similar solvents Do not use abrasive cleaners on any portion of the pow...

Reviews: