Accuracy Table
6
Resistance
Range
Accuracy
Measurement current
600.0
Ω
±0.9% rdg. ±5 dgt.
200
μ
A ±20%
6.000 k
Ω
±0.9% rdg. ±5 dgt.
100
μ
A ±20%
60.00 k
Ω
±0.9% rdg. ±5 dgt.
10
μ
A ±20%
600.0 k
Ω
±0.9% rdg. ±5 dgt.
1
μ
A ±20%
6.000 M
Ω
±0.9% rdg. ±5 dgt.
100 nA ±20%
60.00 M
Ω
±1.5% rdg. ±5 dgt.
10 nA ±20%
• Open circuit voltage: 1.8 V DC or less (DT4221, DT4222)
2.0 V DC or less (DT4223, DT4224)
• Overload protection: 600 V DC/600 V AC or 3 × 10
6
V • Hz (energized for 1 minute)
Current under overload: Steady state 15 mA or less, transient state 0.8 A or less
(DT4221, DT4222)
• Overload protection: 750 V DC/750 V AC or 3 × 10
6
V • Hz (energized for 1 minute)
Current under overload: 1.5 mA or less (DT4223, DT4224)
• Maximum capacity load: 10 mF
• Maximum inductive load: 10 H
• Accuracy guarantee condition: After zero adjustment has been performed
• Auto range movement threshold: 6000 counts or more for upper range.
Switches to the high impedance 60.00 M
Ω
range when the test leads
are open, regardless of the range.
540 counts or less for lower range.
7
Capacitance
Range
Accuracy
Charging current
1.000
μ
F
±1.9% rdg. ±5 dgt.
10 n/100 n/1
μ
A ±20%
10.00
μ
F
±1.9% rdg. ±5 dgt.
100 n/1
μ
/10
μ
A ±20%
100.0
μ
F
±1.9% rdg. ±5 dgt.
1
μ
/10
μ
/100
μ
A ±20%
1.000 mF
±1.9% rdg. ±5 dgt.
10
μ
/100
μ
/200
μ
A ±20%
10.00 mF
±5.0% rdg. ±20 dgt.
100
μ
/200
μ
A ±20%
• Open circuit voltage: 1.8 V DC or less (DT4221, DT4222)
2.0 V DC or less (DT4223, DT4224)
Summary of Contents for DT4200
Page 2: ......
Page 6: ...iv Contents ...
Page 60: ...54 Power on Option Table ...
Page 76: ...Appx 2 RMS and Average ...
Page 77: ...16 01 EN ...
Page 78: ......
Page 79: ......