![Tuttnauer TIVA8-L Instructions For Use Manual Download Page 49](http://html1.mh-extra.com/html/tuttnauer/tiva8-l/tiva8-l_instructions-for-use-manual_1182786049.webp)
49
Operative Instructions
ALARM
DESCRIPTION
ACTION
E70 PUMP PRESSURE
There could be water leaks.
Press RESET
If it persists, call the service
personnel.
E71 HEPA FILTER
The HEPA filter might be dirty or clogged.
Check it.
Press RESET
If it persists, call the service
personnel.
E75 PRODUCT 1 EMPTY
Chemical liquid tank 1 is empty.
Add liquid.
Press RESET
If it persists, call the service
personnel.
E76 PRODUCT 2 EMPTY
Chemical liquid tank 2 is empty.
Add liquid.
Press RESET
If it persists, call the service
personnel.
E77 PRODUCT 3 EMPTY
Chemical liquid tank 3 is empty.
Add liquid.
Press RESET
If it persists, call the service
personnel.
E78 PRODUCT 4 EMPTY
Chemical liquid tank 4 is empty.
Add liquid.
Press RESET
If it persists, call the service
personnel.
E80 SPRAY ARM LOCKED
The spray arm marked in red on display
is locked. Remove it and clean.
Press RESET
If it persists, call the service
personnel.
12.7 Warnings table
WARNING TYPE
DESCRIPTION
ACTION
LIQUID RESERVE
PRODUCT 1
Product 1 is in reserve or may be empty.
Add the product 1
LIQUID RESERVE
PRODUCT 2
Product 2 is in reserve or may be empty.
Add the product 2
LIQUID RESERVE
PRODUCT 3
Product 3 is in reserve or may be empty.
Add the product 3
LIQUID RESERVE
PRODUCT 4
Product 4 is in reserve or may be empty.
Add the product 4
REQUEST
MAINTENANCE
Contact customer service or the dedicated
technical assistance for periodic maintenance.
Press RESET
REQUEST
VALIDATION
Contact customer service or the dedicated
technical assistance for periodic validation.
Press RESET
REFILL
SALT TANK
1- Take out the rack
2- Unscrew the cap of the salt tank
3- Fill the salt tank
4- Screw the cap of the salt tank
Press RESET
For 5 seconds
Summary of Contents for TIVA8-L
Page 55: ...55 Operative Instructions...