![Tuttnauer TIVA8-L Instructions For Use Manual Download Page 37](http://html1.mh-extra.com/html/tuttnauer/tiva8-l/tiva8-l_instructions-for-use-manual_1182786037.webp)
37
Operative Instructions
a
WARNING! It is mandatory to record the
monthly air suction filter cleaning in the
Maintenance Logbook, to ensure continuous
optimal operation.
11.2.4 Replacing the air suction filter
The glasswasher is standard equipped with a suction
filter for drying air.
On request, the machine can be equipped with an
«absolute» «HEPA H14» class supplementary air filter,
in compliance with EN 1822.
a
WARNING! It is mandatory to replace
the prefilter and the HEPA H14 filter as
indicated in the Maintenance Logbook. In
case of intensive use of the machine it is
recommended to increase the frequency of
maintenance.
When replacing the air suction filter use a Torx
screwdriver T20 to open the filter panel.
11.3 Special maintenance
Special maintenance operations are not foreseen by
the user but must be performed exclusively by the
technical assistance or by an authorised and qualified
service personnel.
How to carry out special maintenance operations and
the frequency of maintenance are described in the
Service Manual.
Contact the Manufacturer service for special
maintenance
.
a
WARNING! Special maintenance is carried
out by the technician under guarantee only if
the uilizer has correctly and regularly carried
out all ordinary maintenance operations and
if the uilizer has filled out and kept up-to-
date the Maintenance Logbook
.
11.4 Troubleshooting
The annex 12.8 lists the main malfunctions, with
relative causes and recommended solutions, that the
machine may present during its operation.
The interventions that may become necessary must be
carried out by experienced and qualified operators.
If the problems persist or appear even more often
after performing the work reported below, contact
the technical assistance or an authorised and qualified
service personnel.
11.5 Equipment disposal
Respect the laws in force in the country of use of the
machine, in relation to the use and disposal of the
products used for the cleaning and maintenance of the
machine, as well as observe what the manufacturer of
these products recommends.
When disposing of a glasswasher, it is necessary
to remember that it can still be contaminated by
blood and other organic liquids, pathogenic germs,
genetically modified material, toxic or carcinogenic
substances, heavy metals, etc. and, therefore, it must
be decontaminated before its disposal.
For safety and environmental protection reasons,
dispose of all chemical residues in compliance with
applicable legislation. When performing this operation,
use protective goggles and gloves. Remove or disengage
the door lock so that no one can get closed inside the
machine, e.g. children playing. Finally, deliver the
machine to a suitable and authorised collection centre.
The electrical and electronic devices to be disposed of
contain reusable materials.
They also contain components harmful for the
environment, but necessary for the correct operation
and safety of the equipment. If they are not disposed
of correctly or are disposed of as household waste,
these components can damage human health and the
environment. Never dispose of the old glasswasher
among conventional waste.
Unauthorised product disposal by the user entails
the application of very strict administrative sanctions
provided for by current legislation. Enquire about this
at your local dealer. Depending on the country in which
one resides and the regulations in force, one is obliged
to delete the data referring to people and stored on the
machine.
Finally, make sure that the old equipment is kept out
of reach of children until its actual disposal.
11.6 Spare parts
The various components of the machine can be
requested directly to the manufacturer by providing
the following data:
-
Model, serial number and year of construction
of the machine.
These data are stamped on the
identification plate fitted on each individual machine.
Summary of Contents for TIVA8-L