47
Operative Instructions
12.6 Alarm table
ALARM
DESCRIPTION
ACTION
E00 EEPROM ALARM
Firmware installed successfully.
1- Press RESET
2- Install new dataset
E01 BLACKOUT
A blackout occurred during cycle
execution and the program was stopped. Press RESET
E02 DOOR OPEN
The door is open or unlocked.
Close the door.
Press RESET
If it persists, call the service
personnel.
E07 NO LOCKED DOOR
The door did not lock within the set
time.
Press RESET
If it persists, call the service
personnel.
E09 NO UNLOCKED DOOR
The door did not unlock within the set
time.
Press RESET
If it persists, call the service
personnel.
E11 DIRTY WATER
Check for dirt on the load and check the
inlet demineralized water.
Press RESET
If it persists, call the service
personnel.
E20 NO COLD WATER
The cold water tap could be closed or
partially open. Check it.
Check if the safety floatswitch
has not tripped.
Press RESET
If it persists, call the service
personnel.
E21 NO HOT WATER
The hot water tap could be closed or
partially open. Check it.
Press RESET
If it persists, call the service
personnel.
E22 NO DEMI WATER
The demineralized water tap could be
closed or partially open. Check it.
Press RESET
If it persists, call the service
personnel.
E26 PRINTER ERROR
The printer is turned off, unplugged or
out of paper. Check it.
Press RESET
If it persists, call the service
personnel.
E30 FLOWMETER 1
Chemical liquid tank 1 might be empty.
Check it.
Press RESET
If it persists, call the service
personnel.
E31 FLOWMETER 2
Chemical liquid tank 2 might be empty.
Check it.
Press RESET
If it persists, call the service
personnel.
E32 FLOWMETER 3
Chemical liquid tank 3 might be empty.
Check it.
Press RESET
If it persists, call the service
personnel.