12.2 Chamber Clean cleaning procedure
The Chamber Clean cleaning program for the family of Tuttnauer's T-Edge
Autoclaves is described below:
Important :
To avoid burns when removing the trays and holders, the Chamber must
be cold.
1. All steps in this procedure must be completed without interruption.
2. Open the autoclave chamber door.
3. Remove instruments and any load, trays from the autoclave chamber.
4. Place the trays and tray holder in the sink to be cleaned with a stainless
steel Safe cleaner.
5. Place one tablet (1) from the tablet packing (2) inside the sterilization
chamber at the back bottom of the cold chamber.(see figure below).
6. Close the autoclave door.
7. Fill the clean water reservoir with
distilled water only
. (Either Automatic or
Manual filling) until the screen indicates full.
Page 91