1.10.2
T-Edge11 fit into sterilization center cabinet - recommended
dimensions:
To allow adequate cooling & ventilation of the autoclave placed inside the
sterilization center cabinet, the following T-Edge11 cabinet sizes are
recommended:
Width - 23.65”- 24.40”/60-62cm; Height – 22.25”- 22.65”/56.5-57.5cm; Min.
Depth - 24.9”-25.2”/63.2-64cm
Note:
The imperial dimensions in inches/lbs are approximated to less than 3
percent.
1.10.3
Device Properties
Table 2 - Properties of the device
Property
Dimension
External size
Width
~19.7"(50cm)
Height
~19.7"(50cm)
Depth
~22.8" (58 cm) supporting common install base
carry a ~23" (60 cm) counter top
Chamber
Diameter
~11"(28 cm)
Depth
~18"(46 cm)
Volume
~913 Ounces (27.2lit)
Usable chamber
space
75% ( ~685 Ounces / ~20.5 L)
Max. Allowable Working Pressure
(MAWP)
~40.6PSI (2.8 bar)
Maximum load per item
~0.45 lbs. (0.2kg)
Maximum
load per tray
Unwrapped
~4 lbs. (1.8kg)
Wrapped
~2.4 lbs (1.08kg)
Maximum
Solid load
Unwrapped
~19.8 lbs (9kg)
Wrapped
~11.9 lbs (5.4kg)
Maximum textile load
~4.4 lbs (2kg)
Page 10
Summary of Contents for AMS11-120-T
Page 2: ......
Page 4: ...https www tuttnauerusa com...
Page 16: ...1 10 Device Specifications 1 10 1 Overall Dimensions Page 9...
Page 56: ...Note For the results of the cycle see description in section 8 5 Page 49...
Page 67: ...Press to receive the entire printout of the selected cycle history Page 60...
Page 86: ...Page 79...