Message
Description
cannot reach the required temperature, in
the chamber, within the preset time.
"Low Pressure"
This message is displayed if Chamber
Pressure drops below the sterilization
pressure for 2 seconds during the
sterilization stage.
"High Pressure"
This message is displayed if Chamber
Pressure raises 0.29 bar above
sterilization pressure for 2 seconds during
the sterilization stage.
"High Pressure (Ending)"
This message is displayed if the system
cannot reach atmospheric pressure ±
0.05 bar during the ending stage.
"High Pressure (Exhaust)"
This message is displayed if the system
cannot reach preset pressure within 10
minutes from the beginning of the
exhaust stage.
"Pressure Time Error"
This message is displayed if the system
cannot reach the required pressure
conditions in the chamber, after preset
time, during the air removal stage.
"Time Error"
This message is displayed if the real time
clock is faulty.
"Door is open (During the
cycle)"
This message is displayed when the door
is open: During the cycle.
"Canceled By User"
This message is displayed after the
STOP button is pressed and cycle
aborted.
Page 73