4.3
Program 3 - Wrapped pouches
For wrapped pouches and materials, when the manufacturer recommends
autoclaving at temperatures of 134
C (273
F) with a drying stage.
Nominal parameters default settings
Sterilization temperature: 134
C (273
F)
Sterilization time: 7 minutes.
Drying time: 30 minutes (may be altered by the operator)
Operations Sequence
Air-removal stage; vacuum pulses are performed.
Heating stage: steam is inserted into the chamber until the sterilization
temperature is reached.
Sterilization stage; temperature and pressure are maintained constant at
the pre-set level for the sterilization time.
Fast exhaust stage; steam is exhausted out of the chamber at a fast rate
until pressure decreases to ambient pressure.
Dry stage; heating of chamber followed by a vacuum break (air inlet) to
remove leftover moisture from the instruments and wraps.