Page 22 of 56 Pages
3.3
Displayed operational messages / Symbols
Message /
Symbol
Name
Message / Symbol
Description
Required Action
This symbol is displayed
when the door is open.
Close the door.
Door is open
(during stand
by)
This message is displayed
when the door is opened: In
standby - if START/STOP
is preset.
Close the door to perform a
new cycle.
Cycle Ended
This message is displayed
when the cycle ended
successfully.
Press START/STOP in
order to perform a new cycle.
Test Ended
This message is displayed
when the test ended.
Press START/STOP in
order to perform a new test
This symbol is displayed
when Cycle by Clock mode
is performed.
Enter the Admin menu as
described in this manual to
change the time or to cancel
this option.
This symbol is displayed
when the temperature in the
jacket did not reach the
preset value
Wait until the temperature in
the jacket reaches the preset
value
Cycle by
clock
This message is displayed if
the user presses
START/STOP key while
the "cycle by clock" mode
is active.
Enter the Admin menu as
described in this manual to
change the time or to cancel
this option.
Atmospheric
pressure not
set
This message id displayed
in order to set the
atmosphere pressure by
opening the door for 5
minutes.
Open the door for 5 minutes
in order to set the
Atmospheric pressure.
Critical
settings have
been updated,
Please restart
machine in
order for
changes to be
updated
If a change of the autoclave
setting was made, a restart
operation is required.
Restart the autoclave in
order for changes to be
updated.