Page 12
3.5
Environment Emission Information
A.
Peak sound level generated by the sterilizer is « 70 / dBA with a back
sound level of 60 dBA.
B.
Total heat transmitted by the sterilizer is < 240 W/h
3.6
Electrical Data
3.7
Water quality
Physical Characteristics and Contaminants Levels
The distilled or mineral
–free water supplied to the autoclave should have the
physical characteristics and maximum acceptable level of contaminants
indicated in the table below:
Physical Characteristics and Maximum acceptable contaminants levels
in steam for sterilizers (According to EN 13060: 2010)
Element
Condensate
– allowable content
Silicon dioxide SiO
2
≤0.1 mg/kg
Iron
≤0.1 mg/kg
Cadmium
≤0.005 mg/kg
Lead
≤ 0.05 mg/kg
All other metals except iron,
cadmium, lead
≤0.1 mg/kg
Chloride (Cl)
≤0.1 mg/kg
Phosphate (P
2
O
5
)
≤0.1 mg/kg
Conductivity
≤3 μs/cm
pH value (degree of acidity)
5 to 7
Appearance
Colorless clean without sediment
Hardness (Σ Ions of alkaline earth) ≤0.02 mmol/l
Property
Value
Total Power
2400W (2*1200W)
Voltage
230VAC/1Ph
Amperage
10A
Protection against electrical shock
Class I (IEC 60601-1)
Mains supply fluctuation
+/- 10%
Degree of protection by enclosure