45
C
orrec
ti
on
s
10.1
C
he
ck
se
cti
on
6
for
c
orr
ec
t
ope
ra
ti
ng
proc
edur
es
.
S
ee
se
c. 4 fo
r installation
inst
ruc
ti
ons.
10.2
Ma
ke
sure
no it
ems a
re
t
ouc
hing t
he
wa
ll
s
of the
c
ha
mbe
r.
10.3
S
et
the ste
riliza
ti
on ti
me a
cc
ording
to t
he
table
in sec
ti
on
6
.
10.4
S
ter
il
ize
fe
w
er tow
els or
gowns or
a
djust
the f
ront o
f the
unit
higher
t
o a
ll
ow mor
e
wa
ter
int
o the c
h
amber
.
S
ee
inst
all
ati
on
inst
ruc
ti
ons sec
ti
on 4.2
/m
axim
um l
oa
d
table
se
cti
on 2.
10.5
Tight
en
the door
more
.
If
lea
king per
sist
s,
re
plac
e the door
g
aske
t.
10.6
Door
be
ll
ows is l
ea
king,
ha
ve
a
te
chnicia
n
re
plac
e the be
ll
ows
.
10.7
F
oll
ow
proc
edure
in se
cti
on
7
.5
.
If
lea
king
pe
rsist
s
ha
ve
tec
hni
cian r
eplac
e th
e S
afe
ty
va
lve
.
10.8
Ha
ve
a
t
ec
hnicia
n
che
ck
the ma
chine
.
11.1
F
oll
ow in
struc
ti
on in s
ec
tion
7.
12.1
Apply
1 or 2
drop
s of 3 i
n 1 oil
or some
white lit
hi
um gre
ase
to t
he
sc
re
w a
nd
be
aring of
the
closi
ng de
vice
,
se
e s
ec
7
.
Possi
ble Cau
se
C
h
ec
ku
p
an
d Te
sts
10.1
C
ha
mber
wa
s not
fill
ed
with t
he
c
orr
ec
t
amount
of w
ater
.
10.2
Ite
ms ar
e lyi
ng up
ag
ainst
the c
ha
mber
.
10.3
The
ster
il
iza
ti
on ti
me is set too
high.
.
10.4
Ite
ms
be
ing st
erilize
d a
re
a
bsorbing a
ll
ava
il
able ste
am (
clot
h towe
ls or gowns)
.
10.5
Door
ga
ske
t i
s lea
king s
te
am.
10.6
S
tea
m i
s lea
king at the c
losing
d
evice
.
10.7
S
afe
ty Va
lve
is l
ea
king.
10.8
The
re
is an int
erna
l st
ea
m l
ea
k or e
le
ctr
ica
l
proble
m.
11.1
Door
be
ll
ow loc
king pi
n is t
ra
ppe
d.
12.1
C
losi
ng de
vice
ne
eds oi
l.
S
ympt
om
10.
Ite
ms
in t
he
c
ha
mber
a
re
burning
or me
lt
ing
.
11.
C
losi
ng de
vice
doe
s not
ope
n
whe
n pre
ssur
e ga
ug
e
re
ads 0
psi.
12.
C
losi
ng de
vice
is har
d to
close
.