34
Note:
The sooner the Multi-purpose valve is turned to Exh / Dry at
the end of the sterilization cycle, the more effective and efficient will
be the drying.
•
When the white needle on the pressure gauge has reached
0 psi the door can be opened.
•
Unscrew the Door Closing Devise as far as it will go, but do not move
it to the side, this will allow the door to open about ¾ of an inch.
•
Leave the Multi-purpose valve knob in the Exhaust / Dry position.
•
Reset the Timer for drying, 20 – 30 minutes, the Dry Light will come
on indicating that Drying is active, and the Heating Elements are back
on.
•
When the Timer reaches 0 min., the drying is complete, and the Dry
Light and Heating Elements will turn off.
•
Unscrew the Door Closing Device, move it to the side and open the
door to remove the instruments.
•
Now turn the Multi-purpose valve knob to the “0” or off position.
Warning
Multi-purpose valve knob should be turned in a clockwise direction
only.
4.
At the end of the day, turn the ON / OFF rocker switch to the OFF position.