5.3 Cycle
End
At end of a successful sterilization cycle, a buzzer sounds four times, and the CYCLE
COMPLETE message appears. The Cycle Status indicator will be blank and grey.
Tap the
to release the door locking mechanism.
The door locking mechanism will releases only if these conditions are fulfilled:
The chamber pressure is below 115 kPa,(2.1 psig).
There is no liquid in the chamber.
The chamber temperature is below the END TEMP parameter.
Open the door and remove the sterilized material from the loading cart following the
steps indicated in the ‘Un-loading’ paragraph.
Warnings!
Loading or unloading of material from the chamber can only be done if the door is
completely open and in the lowest position.
It is forbidden to put any part of the body inside the chamber whilst the door is
closing.
It is strictly forbidden for any person to enter the autoclave’s chamber.
If, for any reason (for cleaning, maintenance or if something falls down), it is
necessary to enter the chamber, the person must shut the system by actuating
the emergency switch and withdraw the key from the system, keeping the key on
him, to prevent accidents and injuries.
Do not operate the autoclave in the presence of dangerous gases and vapors.
Warning!
Verify that the pressure in the chamber is atmospheric pressure. Do not attempt
to open the door if the pressure in the chamber is above atmospheric pressure.
Note:
If cycle failed in 2-door autoclave, only the
front door
can be opened.