MAN205-0097044ENO Rev. A
Apr. 2014
Page 28 of 28 pages
Message
Description
Corrective Action
High Pressure
(Ending)
If chamber pressure is above
atmospheric pr 5 kPa X
minutes (as defined in cycle
parameter 'Normal Pressure Time
Out) after the end stage begins
Call for service
Air Error
If chamber pressure is below
atmospheric pressure - 5kpa X
minutes (as defined in cycle
parameter 'Normal Pressure Time
Out) after the end stage begins
Call for service
High Temp. (Ending)
If chamber temperature does not
reach X°c as defined in cycle
parameter 'End Temperature' X
minutes (as defined in cycle
parameter 'Normal Temperature
Time Out) after the end stage
begins
Perform a new cycle
Accessory Timeout
In Accesory test cycle - If Digital
input 'Acc Operate' value is 0 after x
minutes as defined in cycle
parameter 'Accessory Time Out'
Call for service
Emergency Stop
If emergency stop has been pushed
during cyle (If Digital input
'Emergency Stop' value is '0')
Call for service
Bio Filter Time Out
If bio hazard filter is not ready after
2 minutes from the start of stage
'Purge'
Call for service
Purge Time Out
If chamber temperature did is below
X°c as defined in cycle parameter
'Purge Temperature' X minutes (as
defined in cycle parameter 'Purge
Time Out) after the purge stage
begins
Call for service
Air Detector
if value of 'Air Detector Pressure' is
bigger than value defined in cycle
parameter 'Air Detector'
Call for service