iGAS Operating Instructions, Issue 6, Apr 2018
Save iGAS
√
Save the iGAS readings in the dataset
Save iEXTRA
√
Save iEXTRA readings in the dataset
Save iDUST
Save the iDUST readings in the dataset
Save iMET
Save the iMET readings in the dataset
Save housekeep.
Save the Housekeeping readings in the dataset
Save diagnostic
Save the Diagnostic readings in the dataset
Save GPS
Save the GPS co-ordinates if GPS module is fitted
Allow to sleep
√
Allow to fall asleep after about 5 minutes if not sampling
and no communications. Toggle switch or send break to
command from WS3 to awaken. Will automatically awaken
and not fall asleep if +12V connected and battery > 6.5V
Dummy readings
For test purposes only
Calibrate iGAS
Use default 1000nA/ppm sensitivity
Calibrate iDUST
Apply calibration to iDUST module
Use time server
Use internet time server to set time, requires WiFi module
Use WiFI
Use WiFi, requires WiFi module
Measure manifold pressure
Measure manifold flow pressure
Sample T&RH
Sample temperature and humidity sensor is fitted
Power photoionization detector
Supply power to PID, sensor must be fitted
Lamp can be operated intermittently to save power
Measure atmospheric pressure
Sensor is fitted by default. Measures enclosure T too
Measure ambient T & RH
External T & RH sensor must be fitted
Measure ambient CO2
Measure ambient CO2, sensor must be fitted
Measure iDUST sample flow
Measure iDUST sample flow, sensor must be fitted
Measure wind speed & direction
Requires iMET with anemometer module
Measure insolation
Requires iMET with solar module fitted, measures intensity
of sunlight in W/m
2
Measure rainfall
Requires iMET with tipping bucket rain gauge.
INSTRUMENT INFORMATION
(show in
bold font
in the text)
The Instrument Information table (II) provides additional information about the instrument. It is read
only and cannot be modified by the user. Some of the values (SEDs) are automatically uploaded from
the Manifold. Not all of the Instrument Information is listed below.
Information Table (II)
Explanation/Description
Serial Number
Instrument serial number
Instrument type
iGAS Monitor
Manufacturer
Turnkey UK
Software Version
Software version programmed into instrument
Gas Manifold
iGAS Manifold serial number
Cal. Due by
Date next instrument re-calibration is due.
CH0 to CH7 names
Names of the primary measurements
CH0 to CH7 units
Measurement units for the primary measurements
Gas-A to Gas-H cal. dates
Calibration dates of the individual gas sensors
Gas-A to Gas-H type
Species and type of the individual gas sensors. For
example, NO-B4 is a 4 -electrode electrochemical
nitric oxide sensor