Addendum: Clean-In-Place System
E09 TD-4100XD
A-8
E09 TD-4100XD Oil In Water Monitor
Clean-In-Place Addendum
PN 106169
Rev B
Bubble trap valve actuator shown with the
cover removed.
The bubble trap valve and outlet valve
must be wired as shown here. Connect
to the monitor, using glands and cabling
appropriate for your hazardous area
requirements.
Green (ground)
White (neutral)
Black (normally closed)
Red (normally open)
CIP system pump shown in shipping
crate, front panel removed. To make
electrical connections at the CIP pump:
1. Remove the four bolts connecting the
pump to the stainless steel panel.
2. Unfasten the pump from the tank by
loosening the O-ring.
3. Remove the cap at the bottom right
side of the pump to access the wiring.
4. Connect to the monitor, using glands
and cabling appropriate for your
hazardous area requirements.
5. Reverse these steps to return the
pump to its mounted position.
Internal switch housing view
After the CIP tank float switch is installed,
as described in Step 7 of
Installing the Clean-In-Place System
make electrical connections from the
switch to the pump and to the monitor
terminal blocks (I). Use glands and
cabling appropriate for your hazardous
area requirements.
The black, white, and green wires must
be connected as follows: the green leads
can be connected together with a wire nut;
the white leads can also be connected
together with a wire nut; one black lead
should be connected to the normally open
terminal and the other to the common
terminal.
NOTE
: When mounting the switch, the
arrow on the side of the switch must point
up. The conduit port must point down.