Addendum: Clean-In-Place System
E09 TD-4100XD
A-6
E09 TD-4100XD Oil In Water Monitor
Clean-In-Place Addendum
PN 106169
Rev B
The CIP inlet valve actuator shown in a remote
installation. The 1/2 in. CPVC ball valve switches inlet
from sample stream to CIP system tank.
Turner Designs Hydrocarbon instruments strongly
recommends installing a vacuum breaker, such as the
Plast-O-Matic VBM050V-CP, near the valve outlet.
A pressure regulator is shipped with the monitor and
should be installed on the sample water inlet pipe,
before the actuator.
Note the 3° slope in the outlet piping to the monitor.
Vacuum breaker
Valve outlet pipe
Sample water inlet pipe
Cleaning solution inlet pipe
The monitor sample supply connection is described in
Section 2.3.2.2 Installation – Piping Connection
.
The monitor sample return connection is described in
Section 2.3.1.2 Installation – Piping Connection
.
A.1.3 MAKING ELECTRICAL CONNECTIONS
WARNING:
Electrical connections should be made only by trained service personnel or certified
electrician, and should only be completed in accordance with site isolation, safe work practices,
and appropriate electrical codes.
The information in this section replaces the relay assignments described in
Section 5.7 Configuring
Alarms and Relays
.
The following image shows the monitor terminal blocks, located inside the enclosure door on the access
panel.
Figure A.1.3.1 Monitor access panel and terminal blocks