3. Calibration Software and Calculations
The Little Dipper
2 is factory adjusted to the following concentrations:
2820-000-A
200ppb PTSA (18.0mA
– 19.5mA)
2820-000-F
100ppb Fluorescein (18.0mA
– 19.5mA)
If your application requires a different concentration range, the Calibration Cable kit P/N
2820-505 is required. Contact
to purchase. Refer to calibration
instructions below:
Note: Refer to Appendix E if integrating with
a Control Box.
3.1 Software Requirements and PC Installation
The LD2 software user interface has been designed to work with Windows-based systems.
At this time there is no Macintosh version available.
3.1.1 Minimum PC Requirements
Intel
®
Pentium
®
133 MHz microprocessor
Microsoft
®
Windows
®
98 or higher
32 MB of RAM
800 x 600 SVGA resolution
3.1.2 Software and Calibration Cable Installation
Included on the USB flash drive are:
LD2 application
Driver installer for 32-bit Windows OS
Driver installer for 64-bit Windows OS
Release Notes
License Agreement
1. Exit all Windows
programs
2. Plug the USB connector of Calibration Cable into your computer.
3. Insert the
USB Flash Drive (also located on the Turner Designs Customer Care
Software / Firmware webpage under
Little Dipper
2).
4.
If you’re running 64-bit Windows OS, double-click the
CP210xVCPInstaller_x64.exe icon to install the correct driver for your operating
system. If you’re running 32-bit Windows OS, double-click the
CP210xVCPInstaller_x86.exe icon.
Calibration Cable Kit P/N 2820-505
12v Power Supply
Calibration Cable